14
8018429/1DWP/V3-0/2022-06| SICK
A D D E N D U M T O O P E R A T I N G I N S T R U C T I O N S | GMS840
Subject to change without notice
3
INSTALLATION
3
Installation
3.1
Safety during transport and mounting
3.1.1
For GMS841 (ATEX/ IECEx) and GMS842 (
C
CSA
US)
3.2
Scope of delivery
Please see the delivery documents for the scope of delivery.
CAUTION: Risk of injury
▸
Consider the weight of the enclosure before lifting,
see “Enclosure specifications”,
▸
Call in further personnel as assistants as required.
CAUTION: Risk of injury
There is a risk of injuries should the enclosure drop down due to the weight of the device
and through hard, protruding enclosure parts.
During transport and mounting:
▸
Wear safety shoes. Wear non-slip gloves.
▸
Handle the device carefully and safely. Secure during transport. Avoid falls and
collisions.
▸
Call in further personnel as assistants as required.
NOTE: Risk of damage
▸
Do not use gas connections and cable inlets as lifting points.
WARNING: Risk of explosion through sparks
▸
Do not carry out work on the device in potentially explosive atmospheres.
▸
Do not unpack the device in potentially explosive atmospheres.
▸
Only use an adequate packaging for transportation of the device.
▸
Only use tools approved for Ex zones.