A D D E N D U M T O O P E R A T I N G I N S T R U C T I O N S | GMS840
Subject to change without notice
INSTALLATION
3
3.6.3
Connecting the signal connections (as required)
3.6.3.1
I/O module (standard)
The standard version has a built-in I/O module. A second I/O module can be fitted (option).
▸
Position of signal connections,
see “Electrical connections”, page 17
.
▸
Function of signal connections
→
“I/O module” Addendum to Operating Instructions.
▸
The cables have to be approved for the relevant application.
– Only use sufficiently screened cables.
– The shielding braid has to end in the cable duct.
– Shorten the shielding braid accordingly.
Fig. 5: Shielding braid
3.6.3.2
Intrinsically safe signal connections (option)
see “Intrinsically safe signal connections (option)”, page 20
3.6.4
Connecting the interfaces (as required)
To use an interface:
▸
Connect the interface cable to the corresponding interface inside the enclosure,
“Electrical connections”, page 17
Function of interfaces
→
Operating Instructions “GMS800 Series”.