30
8018429/1DWP/V3-0/2022-06| SICK
A D D E N D U M T O O P E R A T I N G I N S T R U C T I O N S | GMS840
Subject to change without notice
8
TECHNICAL DATA
8.2
Enclosure specifications
8.3
Ambient conditions
8.4
Gas connections
Version
Approval conditions
For GMS841 (ATEX/ IECEx)
Design:
Closed steel sheet enclosure
Degree of protection:
IP 66 (Nema 4X)
Dimensions:
Weight:
Depending on equipment: max. approx. 30 kg
Identification for GMS84x:
see “Product identification”, page 9
Atmospheric influences:
Only for use indoors
Allowable contamination:
Degree of contamination 2
[1]
[1] Dry and wet contamination that can be electrically conductive.
Fitting position (allowable offset):
Only install the enclosure horizontally (cover opens
downwards or upwards).
Geographic height at installation location:
Max. 2000 m
Allowable oscillations (amplitude):
[2]
[2] Observe DIN 15267-3, DIN EN 60068-2-26; as well as specifications for built-in Analyzer modules.
0.035 mm in range 5 … 59 Hz
Allowable vibrations (acceleration):
5 m/s
2
in range 59 … 160 Hz
Jolts:
≤
15 g over 11 ms
[3]
[3] Shock test in accordance with DIN EN 60068-2-27.
Relative humidity:
10 … 95%, non-condensing
Ambient temperature during operation:
+5 … +45 °C (41 … 113 °F)
Transport /storage temperature:
–10 … +70 °C (14 … 158 °F)
Description
Material
Suitable for
Inner thread G ¼"
[1]
:
Stainless steel Screw-in screw fittings
Plastic clamping ring screw connection
PVDF
Hose 6x1 mm
Swagelok 6 mm
Stainless steel Metal tube with 6 mm outer Ø
Swagelok ¼“
Stainless steel Metal tube with ¼“ outer Ø
[1] Standard version.
Parameter
Allowable value
Sample gas pressure in sample gas
line:
–500 … +1000 hPa (–0.5 … +1.0 bar)
Sample gas volume flow:
Max. 100 dm
3
/hour
●
Sample gas connection positions,
see “Dimensions, fitting position and
●
Sample gas connections function
→
“GMS800 Series” Operating Instructions.
●
Other sample gas specifications (pressure, volume flow etc.)
→
Addendum to
Operating Instructions for the Analyzer modules fitted.