Kapitel
3
Betriebsanleitung
ISD400
78
© SICK AG • Deutschland • Alle Rechte vorbehalten
8012620/YYZ3/2016-08-01
Mounting
maintained (a >100 mm). This
minimum distance is independent
of the maximum distance between
the two ISDs.
3.2
Mechanical
mounting
Install the devices so the optical
axes of the two devices match at
minimum distance:
–
Line distance
0.2
to 180 m:
Install the two devices facing
one another but rotate one
through 180°.
–
Line distance
3.0
to 180 m:
The two devices can be
installed without 180° rotation.