6
COMMISSIONING
130
8017768/12AX/2019-05-31|SICK
O P E R A T I N G I N S T R U C T I O N S | TIC
Subject to change without notice
The toolbar above the navigation tree offers the following functions in operating mode:
Symbol
Meaning
Load and start site configuration
Loads the configuration file containing the key system parameters into TEMS
Manager and starts the configuration.
If a configuration is loaded while another configuration is running, the system will
automatically stop the current configuration. The new configuration is loaded and
started. System operation is paused during this process.
The function requires login as a user with
Authorized User
authorization or higher.
Stop current site configuration
Stops the configuration that is currently running in measuring mode.
The function requires login as a user with
Authorized User
authorization or higher.
Edit site configuration
Stops the configuration that is currently running and switches the system to editing
mode.
Once in editing mode, you can change the system parameters.
The function requires login as a user with
Authorized User
authorization or higher.
Activate/deactivate expert mode
Switches the export mode on and off.
Expert mode displays all system functions in the navigation area and functions for
editing the parameter values, etc.
Save current site configuration on hard drive
Saves the current site configuration onto a data card. The configuration that is
saved in the file can help SICK Support to resolve fault situations quickly.
The function is available for all authorizations.