28
8015383/ZUQ7/V1-3/2020-06 | SICK
O P E R A T I N G I N S T R U C T I O N S | TRANSIC151LP
Subject to change without notice
3
INSTALLATION
3.5.1
Connecting the TRANSIC151LP transmitter
Fig. 11: TRANSIC151LP - electrical connections
Replacing the TRANSIC151LP cable
1
Ensure the device is disconnected from the mains and potential-free.
2
Remove the cover on the front side of the TRANSIC151LP.
3
Push the cable through the cable gland. (Max. cable length: 5 m).
4
Connect to voltage supply terminals U-in(4)PK (pink) and GY (gray).
5
Connect to analog output I-out (5) YE (yellow) and GN (green).
6
Connect to digital output D-out (6) BN (brown) and WH (white). A NAMUR digital output is
located between terminals BN and WH,
see “Pressure suitability”, page 78
7
Connect the shielding to the cable gland.
8
Close the cable gland. Tightening torque: 10 Nm
9
Ensure the cable gland seals the cable.
10
Close the device front panel.
11
Ensure the enclosure is closed tight.
1
4
5
6
3
2
Maintenance interface (SICK Service Adapter required)
LED lamp
Keypad
Voltage supply
Analog output
Digital output (NAMUR)
Cable colors are oriented to DIN47100.