37
8015383/ZUQ7/V1-3/2020-06 | SICK
O P E R A T I N G I N S T R U C T I O N S | TRANSIC151LP
Subject to change without notice
OPERATION
4
4.4.6
Entering the password (PAS)
1
Entering the password makes additional menu items available.
2
These additional menu items remain accessible for 30 minutes.
3
Observe the safety instructions,
see “Safety information on using the password:”,
Fig. 21: Entering the password
4.5
Menu navigation with password authorization
The maintenance level is open for all interfaces after the password has been entered.
4.5.1
Process pressure: Display and settings (APP)
1
Enter the average pressure in the sample gas. Further information,
.
Adjustable range: 800 … 1400 mbar.
Fig. 22: Reading out and changing the process pressure
4.5.2
H
2
0 content in the process gas (H2O)
1
Enter the average H
2
O value in the sample gas. Further information,
of ambient parameters”, page 40
.
Adjustable range: 0 … 600 g/m
3
.
Fig. 23: Humidity in process gas setting
Menu navigation starts from the top again after the password is entered (Measured
value display).
<
<
<
<
<
<
...
Ent
Ent
Up
Up
Ent
Ent
Back
Back
k
Dn
Dn
NOTE:
●
After entering the password via the keypad, it is recommended to return to the oxygen
statistics display after completing password protected functions.
●
Even though the password expires in 30 minutes, maintenance functions remain
available until you return to the basic functions in the menu structure. A message is
not sent via the keypad when the password expires.
<
<
...
<
<
(0.800 ... 1.400 bar)
Back
Back
k
Ent
Ent
Up
Up
Dn
Ent
Ent
<
...
(0 ... 600 g/m³)
<
<
<
<
<
Back
Back
k
Ent
Ent
Up
Up
Dn
Dn
Ent
Ent