MAINTENANCE
7.1.2
Cleaning the optical components
Check the signal strength
●
Maintenance of the TRANSIC151LP
●
Maintenance warning
●
Error signal that indicates excessive light loss in the sensor.
Inquiry via keypad,
see “Signal intensity(SIL)”, page 36
Using solvents to clean the optical components
When using solvents to clean the optical components, ensure the solvent used is
compatible with the sealing material of the sensors.
Cleaning the mirror and lens
Fig. 39: Location of the mirror in the oxygen measuring probe
WARNING:
Corrosive components can endanger operation
▸
Check all parts, especially those made of stainless steel, for corrosion and exchange
these parts as necessary.
Corrosion on device parts can impair safety with regard to explosion protection, leak
tightness and pressure.
NOTE:
Risk of damage to the TRANSIC151LP through dust or humidity
▸
Only open the TRANSIC151LP in a dry environment free from dust.
CAUTION:
Plant-specific hazards during maintenance work
▸
Observe local regulations with regard to plant-specific protective equipment during
maintenance work.
NOTE:
SICK recommends cleaning the optical components when the signal
strength is below 80%.