O P E R A T I N G I N S T R U C T I O N S | TRANSIC151LP
Subject to change without notice
2
PRODUCT DESCRIPTION
2.3.3
Variant for ambient gas measurement
Fig. 4: TRANSIC151LP-B for ambient gas measurement and wall bracket
When configuring ambient gas measurements, the complete TRANSIC151LP must be
exposed to the O
2
concentration to be measured because the transmitter is part of the O
2
measurement.
The TRANSIC151LP variant for ambient measurements measures oxygen concentrations
from 2 ... 21 vol% O
2
.
The measuring function is lost for oxygen concentrations under 2 vol%.
Non-ex zone
TSA151 voltage supply unit
TRANSIC151LP
Measuring environment
TRANSIC151LP complete in differing O
2
concentrations
Zone: 1/21
0
2
: 2 … 21 vol%
T:
−
20 … +60 °C
p: 0.8 … 1.1 bar(a)
DC 24 V
PELV
Connection TSA151 with TRANSIC151LP
Ex ib M20 x 1.5
3 x 2 x 0.5 mm
2
twisted pair, shielded
d: 7 … 12 mm
L: 0.5 … 5 m (longer lines not allowed)
Shield laid on both sides of the EMC cable gland on the
enclosure.
Analog/digital
output
Ex ib M20 x 1.5
2 x 2 x 0.5 mm
2
twisted pair,
shielded
d: 7 … 12 mm
Voltage:
DC 24 V PELV
Um: 60V
Ex e M20 x 1.5
2 x 1.5 mm
2
2 x 1.0 mm
2
Depending on
line length.
d: 5 … 9 mm
WARNING:
Ex certification becomes void for unallowed wiring
▸
Comply with wiring requirements,