ADJUSTMENT
6
6.1.1
Setting up the gas supply for calibration and adjustment
The TRANSIC151LP can be calibrated and adjusted with ambient air or bottled gas.
6.1.1.1
Using ambient air
See
for further information on this calibration method.
6.1.1.2
Using bottled gas and with sample gas cell
1
Ensure the O-ring is seated securely in the groove.
2
Insert the probe into the sample gas cell.
3
Push the probe against the sample gas cell and turn it clockwise 45°, see
the TRANSIC151LP probe in the sample gas cell”
4
The gas inlets of the sample gas cell are fitted with Swagelok gas connections for
Ø 6 mm tubes,
see “Installing the sample gas line”, page 24
. A 6 mm
↔
1/4‘‘ adapter is
delivered.
5
Let the gas flow out without hindrance. This avoids an overpressure in the chamber.
Fig. 33: Fastening the TRANSIC151LP probe in the sample gas cell