50
8015383/ZUQ7/V1-3/2020-06 | SICK
O P E R A T I N G I N S T R U C T I O N S | TRANSIC151LP
Subject to change without notice
6
ADJUSTMENT
6.3
Adjustment
6.3.1
Adjustment process
1
Enter the password,
see “Entering the password (PAS)”, page 37
2
Access to the adjustment functions is open for 30 minutes after the password is
entered. Functions in progress are not interrupted after expiry of 30 minutes. Enter the
password again to execute more password-protected functions.
3
Make sure that no error messages are active as these could affect adjustment.
Malfunction messages,
see “Display of active and undeleted errors (ERR)”, page 36
.
4
Make sure that the ambient parameters of the adjustment environment have been set
before adjustment.
5
Set the values for pressure, humidity and CO
2
concentrations of the calibration gas.
Calibration gases have a humidity of 0 g/m
3
H
2
O. The CO
2
concentration of nitrogen gas
mixtures is
‐
0 vol% CO
2
.
6
Reset the ambient parameters after adjustment to the values of the process gas.
Further information on the compensation of ambient parameters can be found in
pensation of ambient parameters”, page 40
.
6.3.2
Adjustment options
●
One-point adjustment using the keypad
●
Two-point adjustment using the keypad
●
Restoring factory calibration
6.3.3
One-point adjustment using the keypad (function CAL1)
1
Check that no error messages are active.
Active error messages influence adjustment. Malfunction messages,
active and undeleted errors (ERR)”, page 36
(keypad). Error Table,
.
2
Enter the password in menu
PAS
see “Entering the password (PAS)”, page 37
3
Select menu item
Cal1
. This freezes the analog output.
4
Connect the reference gas.
5
Enter the known O
2
value and confirm with
Ent
.
6
The measured value display blinks.
7
Feed calibration gas.
8
Wait until the display shows a stable value.
9
Confirm with
Ent
.
PASS
is displayed when the calibration is successful. The TRANSIC151LP now calculates
the new gain and offset settings and starts to display the new measured value.
10
Press
Ent
twice.
This ends the one-point adjustment.
●
The reference concentration used determines whether the gain or offset parameter
value is changed.
– Change to the offset value: Oxygen concentration < 10.5% O
2
– Change to the gain value: Oxygen concentration > 10.5% O
2
●
Two-point adjustment: Always returns new gain and offset values.