MAINTENANCE
7.1.3
Checking the temperature probe
Fig. 41: Temperature probe
▸
Check the condition of the temperature probe during every regular maintenance.
▸
Exchange the temperature probe when corrosion can be seen. Contact SICK Customer
Service.
7.1.4
Cleaning the TRANSIC151LP filter
0511-056
Fig. 42: Stainless steel mesh and PTFE filters
WARNING:
Risk of explosion through damaged temperature probe
The temperature probe is part of the zone separation. The partition wall thickness of the
temperature probe is 0.2<t<1 mm.
▸
Ensure the temperature probe is not damaged by corrosive gas.
1
Temperature probe
CAUTION:
Control the filter regularly.
▸
Check the filter regularly.
▸
Change the filter when clogged.
WARNING:
The filter can contain acidic or toxic substances
▸
Observe relevant safety regulations.
▸
Depending on its components, the filter must be disposed of according to legal
regulations, or as hazardous waste when necessary.
Stainless steel mesh filter
PTFE filter