O P E R A T I N G I N S T R U C T I O N S | TRANSIC Extractive
Subject to change without notice
OPERATION
5
5
Operation
5.1
Interfaces
●
Keypad
●
Maintenance interface
5.1.1
Keypad
A display and four pushbuttons are located on the front enclosure panel.
The measured oxygen is shown on the display. LEDs signal the operating mode.
Characteristics
Calibration of:
●
Process pressure
●
Humidity
●
Carbon dioxide content
Fig. 4: Keypad (source: TRANSIC151LP operating instructions)
Display modes
Note
Changes to parameters are password protected. The password allows access
for 30 minutes after entry.
Display mode
Display/LED
Process
Start
(duration: 2,5 minutes)
Software Version
Self test
Pass
Self-test starts
Information: Self-test running
Warming up phase starts
Normal operation
Green LED remains on
Measured oxygen value
Measured oxygen value is
shown continuously
Error state
Red LED remains on
Error state number
Warning
Yellow LED flashes
Measured oxygen value is dis-
played
Select function “Err“ in the
menu