POLYMOBIL Plus
SPR8-125.034.02
Page 2 of 10
Siemens AG
Rev. 02
03.05
CS PS 24
Medical Solutions
1 - 2
Requirements
Tools and measurement devices required
1
•
Standard service kit *
•
Protective conductor test meter
•
Device leakage current meter
•
Digital multimeter.
Safety information and protective measures
1
All tools, measurement and auxiliary devices, with the exception
of those marked "* ", are listed in the STC (Service Tools Cata-
logue) along with their specifications.
Danger of injuries, death or material damage.
Observe
- the product-specific safety information in these instructions,
- the general safety information in the document
TD00-000.860.01...
and
- the safety information according to ARTD Part 2.
Non-compliance can lead to death, to injuries or to material dam-
age.
X-ray radiation!
Non-compliance can lead to illnesses, irreversible damage to
body cells and to the genetic make-up, sever injuries, up to death.
In work on the system in which radiation has to be released, the
radiation protection directives and the rules for radiation protec-
tion according to ARTD-002.731.02 must be complied with.
Observe the following:
- Use available radiation protection devices.
- Wear radiation protection clothing (lead apron).
- Keep distance from the radiation source as large as possible.
- Release radiation only if necessary.
- Set radiation activity as low as possible.
(low KV and mAs values)
- Release radiation as short as necessary.
Checks in which radiation has to be released are identified by the
radiation protection warning symbol.
NOTE
WARNING
WARNING