Commissioning
6.5 Quick Start Wizards
HydroRanger 200 HMI
Operating Instructions, 06/2018, A5E36281317-AC
65
Response rate
Sets the reaction speed of the device to measurement changes in the target range.
Note
•
Any changes made to Fill rate/minute (2.3.2.) (Page 173), Empty rate/minute (2.3.3.)
(Page 173), or Response rate (2.3.4.) (Page 174) parameters following the completion of
the wizard will supersede the Response Rate setting.
•
Response Rate always displays in meters per minute (m/min).
Index
Transducer
Options
Slow (0.1 m/min)
*Medium (1.0 m/min)
Fast (10.0 m/min)
Factory setting
Medium (1.0 m/min)
Use a setting just faster than the maximum filling or emptying rate (whichever is greater).
Slower settings provide higher accuracy, faster settings allow for more rapid level
fluctuations.
Configure another measurement point
Gives option to configure more measurement points or end the wizard
Note
This is available on the dual-point model only.
Options
Yes
Returns to the Measurement Point menu
No
Completes the Quick Start Level configuration
End of QS Level Wizard
Note
For QS to be successful, all changes must be applied.
Options
BACK, CANCEL, FINISH
Display returns to Quick Start (1.1.) menu when Quick Start is successfully com-
pleted or cancelled. If CANCEL is selected, no changes are written to the device.
To transfer Quick Start values to the device and return to PROGRAM menu, press DOWN
arrow (Finish). Then press LEFT arrow three times to return to Measurement mode.