Parameter reference
8.6 Setup (2.)
HydroRanger 200 HMI
Operating Instructions, 06/2018, A5E36281317-AC
165
8.6.1.6
Material (2.1.6.)
Specifies material type.
Index
Single-point model
Dual-point model
Global
Transducer
Options
*Liquid
Solid
Alters
TVT type (2.11.3.6.) (Page 230)
8.6.1.7
Short shot frequency (2.1.7.)
Adjusts the short shot transmit pulse frequency, in kHz.
Index
Single-point model
Dual-point model
Global
Transducer
Values
Range: 41 to 46 kHz, nearest acceptable value is returned
Altered by
Transducer (2.1.5.) (Page 164)
Related
•
Number of short shots (2.1.11.) (Page 166)
•
Long shot duration (2.1.10.) (Page 166)
•
Short shot bias (2.11.2.10.) (Page 225)
•
Short shot floor (2.11.2.11.) (Page 226)
•
Short shot range (2.11.2.12.) (Page 226)
8.6.1.8
Long shot frequency (2.1.8.)
Adjusts the long shot transmit pulse frequency, in kHz.
Index
Single-point model
Dual-point model
Global
Transducer
Values
Range: 41 to 46 kHz, nearest acceptable value is returned
Altered by
Transducer (2.1.5.) (Page 164)
Related
•
Number of long shots (2.1.12.) (Page 167)
•
Short shot frequency (2.1.7.) (Page 165)
•