1 General information
MAMMOMAT Balance
SPB7-115.831.01 Page 5 of 22
Siemens AG
Maintenance Instructions
Rev.01 05.04
Medical Solutions
1 General information
1.1 Training
•
Self-instruction packet or training workshop and at least 1 installation
1.2 Required documents
•
Planning Guide
SPB7-115.891.01.01.02
•
MAMMOMAT Balance Wiring Diagrams
SPB7-115.844.01.01.02
•
Installation and Start-up Instructions
SPB7-115.814.01.01.02
•
Maintenance protocol
SPB7-115.832.01.01.02
•
Instructions for use
SPB7-115.629.01.01.02
•
Service Instruction
SPB7-115.840.01.01.02
1.3 Required tools, measurement and auxiliary devices
•
Oscilloscope
•
Oscilloscope probe 1:100 > 600Vdc
•
Digital multimeter
•
Non invasive kVp-meter (mammo)
•
Dosemeter (mammo)
•
Service PC with serial port (connection cable is supplied with the unit)
•
MAMMOMAT Balance SSW (Service Software Program is supplied with the unit)
•
SIB phantom or Normi 7 or 4 cm Plexi or customer phantom (supplied by customer)
•
Sensitometer / Densitometer
•
Film pack 18x24 and 24x30 if it occurs
•
Cleaning agents specified in the operating instructions
•
Protective conductor meter
•
Bathroom balance with tennis ball
•
Spray enamel according to the PG (to touch up chipped paint)
•
Standard installation tools
•
Plexiglas phantom 2cm, 2cm, 2cm, 1cm, 0.5cm for AEC calibration (supplied with the unit)
1.4 Required lubricants
•
All purpose grease PD2
•
Viscogen oil