1 - 1
Siemens AG
RX57-029.061.01
Page 1 of 4 MEMOSKOP 2/3/50/100/SUB
Medical Engineering
Rev. 02
10.97
TD SD 34
Safety notes
Always switch the system off when performing work on the Memoskop.
Documentation required
1
This document RX57-029.061.01..applies to SIREMOBIL 2000, LITHOSTAR Multiline
and UROSKOP D1/D2.
MEMOSKOP
1
SIREMOBIL 2000
1
LITHOSTAR Multiline
1
Uroskop D1/D2
1
Measurement instruments, auxiliary devices and tools required
1
Protective conductor test
1
After finishing all work steps and attaching all covers, perform the protective conductor
test according to ARTD-002.731.16.
The protective conductor resistance must not exceed 0.2
O
.
•
When working on parts connected to line voltage, there is the risk of electric shock.
•
To comply with EMC regulations, PC board shielding must be installed properly.
•
Always switch the system off or disconnect the power plug prior to removing the covers.
•
MEMOSKOP 2
X2146
•
MEMOSKOP 3/50/(100) and SUB 2000
X2009
•
Service instructions
RR2-120.061.02..
•
Setting the software parameters
RXL2-120.032.02..
•
Service software control
RXL2-120.113.01..
•
Handling service software
RLL5-310.113.02..
•
Calling up the set-up mode (see page 5-1)
•
Digital multimeter Fluke 8060A
97 02 101 Y4290
•
Service PC
•
Service tool kit
•
MEMOSKOP keyboard
11 02 669 X2009
•
Protective conductor meter
44 15 899 RV090
•
Serial PC connection cable for service PC
99 00 440 RE999
Requirements
1