MEMOSKOP 2/3/50/100/SUB
RX57-029.061.01
Page 2 of 4
Siemens AG
Rev. 02
10.97
TD SD 34
Medical Engineering
1 - 2
Requirements
Removing the covers
1
SIREMOBIL 2000
1
Basic unit back cover
•
Lock the foot brake.
•
Switch off the system and disconnect it from the line voltage.
•
Using the Allen key, remove the three screws in the rubber strip (1/Fig. 1).
•
Hold the cover at the bottom and pull it back.
•
Hold the cover at the bottom and pull it downward.
•
Holding the cover at the top, pull up and to the back to remove it.
•
Remove the protective conductor and retighten the protective conductor screw slightly.
Basic unit side cover
•
First remove the rear cover.
•
Remove the screws from the rubber strip (2/Fig.1).
•
Loosen the screws on the top (3/Fig.1).
•
Remove the cover toward the side.
•
Remove the protective conductor and retighten the screw slightly.
Rear cover of the SM2000 monitor trolley
•
Remove the cover screws.
•
Remove the protective conductor.
Fig. 1
3
2
1