Page 2
Heated Enclosure – INSTRUCTION MANUAL
7ML1998IFY01
mmmm
m
Specificat
ions
Specifications
Power
• 115 Vac, 50/60Hz, 120 VA
Environmental
• indoor/outdoor
• altitude: 2000m max.
• ambient: – 40
°
to 50
°
C (–38
°
to 122
°
F)
• relative humidity: suitable for outdoors (Type 4X/NEMA 4X enclosure)
• installation category: II
• pollution degree: 4
• corrosion resistant
Thermostat
• opens at 14
°
C (57
°
F)
• closes
at
7
°
C (45
°
F)
Vertical Enclosure Dimensions
• W. 420 mm (16.5")
• H: 370 mm (14.5")
• D: 215 mm (8.46")
Horizontal Enclosure Dimensions
• W: 370 mm (14.5")
• H: 420 mm (16.5")
• D: 215 mm (8.46")
Important:
• All specifications are subject to change without notice. Please ensure that any
safety related information is confirmed with a qualified Siemens Milltronics
representative.
• Approval information applies to Heated Enclosure only. Please refer to level
instrument documentation for approvals information.
Note:
Thermostat open and closed temperatures have a variability factor
of +/– 5
°
(C/F)
Heater Box.fm Page 2 Wednesday, April 24, 2002 9:47 AM