MOBILETT Plus HP
SPR8-220.061.03
Page 4 of 28
Siemens AG
Rev. 03
09.03
CS PS 24
Medical Solutions
10 - 4
Mechanical systems
Adjusting the friction linings
10
Upper arm
If the upper arm does not hold the position set, the friction lining can be adjusted through
the holes in the cover (Fig. 4).
•
Loosen the screws completely with a 4 mm Allen wrench.
•
Retighten the screws until there is no play.
•
Tighten the screws evenly until the upper arm holds its position but is still easy to move.
Lower arm
If the lower arm does not hold its position, it must be adjusted (Fig. 5).
•
Loosen the screw completely.
•
Retighten the screw until there is no play.
•
Tighten the screw until the lower arm holds its position but is still easy to move.
•
Reattach the cover.