Model Designation and Specifications
UM353-1
March
2003
14-8
Accuracy .........................................+/-0.6°C
Conformity......................................<=
0.06°
Software
Output
Type.....................Analog (configurable
°
C,
°
F,
°
R,
°
K)
Ambient Temperature Effect: ........+/- 0.10°C/°C
Type ‘T’ Thermocouple:
Range Limits...................................-240°C to 370°C (-400
°
F to 698°F)
Performance Range.........................-100 to 370°C
Accuracy .........................................+/-0.5°C
Conformity......................................<=
0.06°
Software
Output
Type.....................Analog (configurable
°
C,
°
F,
°
R,
°
K)
Ambient Temperature Effect: ........+/- 0.07°C/°C
Type ‘E’ Thermocouple:
Range Limits...................................-185°C to 1000°C (-300
°
F to 1830°F)
Performance Range.........................0 to 1000°C
Accuracy .........................................+/-0.5°C
Conformity......................................<=
0.06°
Software
Output
Type.....................Analog (configurable
°
C,
°
F,
°
R,
°
K)
Ambient Temperature Effect: ........+/- 0.07°C/°C
Type ‘S’ Thermocouple:
Range Limits...................................-18°C to 1650°C (0
°
F to 3000°F)
Performance Range.........................200 to 1650°C
Accuracy .........................................+/-0.7°C
Conformity......................................<=
0.06°
Software
Output
Type.....................Analog (configurable
°
C,
°
F,
°
R,
°
K)
Ambient Temperature Effect: ........+/- 0.14°C/°C
Type ‘R’ Thermocouple:
Range Limits...................................-18°C to 1610°C (0
°
F to 2930°F)
Performance Range.........................200 to 1610°C
Accuracy .........................................+/-0.7°C
Conformity......................................<=
0.06°
Software
Output
Type.....................Analog (configurable
°
C,
°
F,
°
R,
°
K)
Ambient Temperature Effect: ........+/- 0.15°C/°C
Type ‘B’ Thermocouple:
Range Limits...................................-18°C to 1815°C (0
°
F to 3300°F)
Performance Range.........................800 to 1815°C
Accuracy .........................................+/-0.7°C
Conformity......................................<=
0.06°
Software
Output
Type.....................Analog (configurable
°
C,
°
F,
°
R,
°
K)
Ambient Temperature Effect: ........+/- 0.15°C/°C
Type ‘N’ Thermocouple:
Range Limits...................................-200°C to 1300°C (-325
°
F to 2370°F)
Performance Range.........................0 to 1300°C
Accuracy .........................................+/-0.5°C
Conformity......................................<=
0.06°
Software
Output
Type.....................Analog (configurable
°
C,
°
F,
°
R,
°
K)
Ambient Temperature Effect: ........+/- 0.10°C/°C
Type DIN 43760/ IEC 751 RTD (à = 0.003850):
Range Limits...................................-185°C to 622°C (-300
°
F to 1152°F)
Accuracy .........................................+/-0.4
°C
Software
Output
Type.....................Analog (configurable
°
C,
°
F,
°
R,
°
K)
Ambient Temperature Effect: ........+/- 0.04°C/°C
StockCheck.com
Summary of Contents for Moore 353
Page 2: ...S t o c k C h e c k c o m ...
Page 14: ...Contents UM353 1 xii March 2003 S t o c k C h e c k c o m ...
Page 24: ...Introduction UM353 1 March 2003 1 10 S t o c k C h e c k c o m ...
Page 152: ...LonWorks Communications UM353 1 March 2003 5 4 S t o c k C h e c k c o m ...
Page 164: ...Network Communications UM353 1 6 12 March 2003 S t o c k C h e c k c o m ...
Page 246: ...Operation UM353 1 March 2003 9 8 S t o c k C h e c k c o m ...
Page 254: ...Controller and System Test UM353 1 March 2003 10 8 S t o c k C h e c k c o m ...
Page 282: ...Circuit Description UM353 1 March 2003 13 4 S t o c k C h e c k c o m ...
Page 298: ...Model Designation and Specifications UM353 1 March 2003 14 16 S t o c k C h e c k c o m ...
Page 302: ...Abbreviations And Acronyms UM353 1 15 2 March 2003 S t o c k C h e c k c o m ...
Page 304: ...Warranty UM353 1 W 2 March 2003 S t o c k C h e c k c o m ...