SITRANS P220 (7MF1567)
6
A5E03304437-AA, 09/2017
WARNING
Use in hazardous areas
Risk of explosion.
•
Only use equipment that is approved for use in the intended hazardous area and labelled accordingly.
See also
Technical specifications (Page 14)
WARNING
Loss of safety of device with type of protection "Intrinsic safety Ex i"
If the device has already been operated in non-intrinsically safe circuits or the electrical specifications have not been
observed, the safety of the device is no longer ensured for use in hazardous areas. There is a risk of explosion.
•
Connect the device with type of protection "Intrinsic safety" solely to an intrinsically safe circuit.
•
Observe the specifications for the electrical data on the certificate and/or in Technical specifications (Page 14).
3
Description
3.1
Application
The pressure transmitter P220, Type 7MF1567, is used for measuring the relative pressure and absolute pressure of gases
and liquids in the following industrial areas:
●
Mechanical engineering
●
Power engineering
●
Water supply
●
Shipbuilding
●
Chemical industry
●
Pharmaceuticals
3.2
Hardware configuration
7MF1567 with plug complying with EN 175301-803-A
•
Type 7MF1567-*****-1**1
•
Type 7MF1567-*****-5**1
7MF1567 with plug M12x1
•
Type 7MF1567-*****-2**1