ATEX/IECEx and others SITRANS TH/TR
SITRANS TF
Display
ГОСТ 31610.0-2014
(IEC 60079-0:2011)
ГОСТ
31610.1.1-2012 (IEC
60079-1:2002)
ГОСТ xxxxx.xx-xxxx/
IEC 60079-31:2013
-
№ TC RU C–
DE.xxxx.x.xxxxxОС НАНИО
«ЦСВЭ»
-
● "Intrinsic safety ia/ib"
type of protection
Zones 0, 1, 2, 20, 21, 22 and M1
Name
● ATEX
II 1 G Ex ia IIC T6 ... T4 Ga
II 2(1) G Ex ib [ia Ga] IIC T6 ... T4 Gb
II 1 D Ex ia IIIC Da
I M1 Ex ia I Ma
II 1 G Ex ia IIC T6 ... T4 Ga
II 1 D Ex ia IIIC Da
I M1 Ex ia I Ma
● IECEx and others Ex ia IIC T6 ... T4 Ga
Ex ib [ia Ga] IIC T6 ... T4 Gb
Ex ia IIIC Da
Ex ia I Ma
Ex ia IIC T6 ... T4 Ga
Ex ia IIIC Da
Ex ia I Ma
● EACEx
0Ex ia IIC T6 ... T4 Ga Х
1Ex ib [ia Ga] IIC T6 ... T4 Gb Х
Ex ia IIIC Da Х
PO Ex ia I Ma Х
0Ex ia IIC T6 ... T4 Ga Х
Ex ia IIIC Da Х
PO Ex ia I Ma Х
Output terminals
SITRANS TH/SI‐
TRANS TF single
chamber housing: 1, 2
SITRANS TR: 3, 4
SITRANS TF dual
chamber housing: (+)
and (-)
U
i
= 30 V DC, I
i
= 120 mA, L
i
=
0 μH, C
i
= 1.0 nF
U
i
= 30 V DC, I
i
= 120 mA, L
i
= ? μH, C
i
= ? nF
U
i
= 30 V DC, I
i
= 120 mA, L
i
= ? μH, C
i
= ? nF
● Permissible
ambient
temperature P
i
=
900 mW
-50 … +85 °C (-58 … +185 °F)
temperature class T4
-50 … +65 °C (-58 … +149 °F)
temperature class T5
-50 … +50 °C (-58 … +122 °F)
temperature class T6
-50 … +85 °C (-58 … +185 °F)
temperature class T4
-50 … +65 °C (-58 … +149 °F)
temperature class T5
-50 … +50 °C (-58 … +122 °F)
temperature class T6
-50 … +85 °C (-58 … +185 °F)
temperature class T4
-50 … +65 °C (-58 … +149 °F)
temperature class T5
-50 … +50 °C (-58 … +122 °F)
temperature class T6
● Permissible
ambient
temperature P
i
=
750 mW
-50 … +85 °C (-58 … +185 °F)
temperature class T4
-50 … +70 °C (-58 … +158 °F)
temperature class T5
-50 … +55 °C (-58 … +131 °F)
temperature class T6
-50 … +85 °C (-58 … +185 °F)
temperature class
T4-50 … +70 °C
(-58 … +158 °F) temperature
class T5-50 … +55 °C
(-58 … +131 °F) temperature
class T6
-50 … +85 °C (-58 … +185 °F)
temperature class T4
-50 … +70 °C (-58 … +158 °F)
temperature class T5
-50 … +55 °C (-58 … +131 °F)
temperature class T6
● Permissible
ambient
temperature P
i
=
610 mW
-50 … +85 °C (-58 … +185 °F)
temperature class T4
-50 … +75 °C (-58 … +167 °F)
temperature class T5
-50 … +60 °C (-58 … +140 °F)
temperature class T6
-50 … +85 °C (-58 … +185 °F)
temperature class T4-50 …
+75 °C (-58 … +167 °F) tem‐
perature class T5
-50 … +60 °C (-58 … +140 °F)
temperature class T6
-50 … +85 °C (-58 … +185 °F)
temperature class T4
-50 … +75 °C (-58 … +167 °F)
temperature class T5
-50 … +60 °C (-58 … +140 °F)
temperature class T6
Technical specifications
7.4 Certificates and approvals
SITRANS TH320/TH420/TR320/TR420/TF320/TF420
68
Compact Operating Instructions, 03/2018, A5E41865021-AA