Scale parameters and functions
8.21 DR 32 display of data and operator errors
SIWAREX WP231
120
Manual, 07/2014, A5E31238908A-02
8.21
DR 32 display of data and operator errors
8.21.1
Overview
Data record DR 32 is used for Modbus communication with a Modbus master. If a function
which is used to write to the holding register is completed with an error, the data or operator
error reported can be read from data record DR 32. Messages are displayed for at least five
seconds and do not need to be acknowledged in the SIWAREX module.
Data record DR 32 does not need to be polled upon positive completion of a function for
writing to the SIWAREX register.
Table 8- 22 Assignment of data record 32
Variable
Note
Type
Length
(bytes)
Rw
Default Min.
Max.
Modbus
registers
Data record
number
Contains no. of data record
USHORT
2
r
32
-
-
3500
Length
Data record length information
USHORT
2
r
28
-
-
3501
Application
Information about which application the
data record belongs to
USHORT
2
r
101
-
-
3502
Version identi-
fier
Information about current data record
version
USHORT
2
r
1
1
65635 3503
5000
Data or operating error exists
BIT
r
0
0
1
3504.16
5001
Command code or data record unknown
BIT
r
0
0
1
3504.15
5002
Command or data change not possible
because write protection is active
BIT
r
0
0
1
3504.14
5003
Leaving calibration mode not possible
BIT
r
0
0
1
3504.13
5004
Command or data transmission only
available in service mode
BIT
r
0
0
1
3504.12
5005
Command or data transmission not pos-
sible because service mode is active
BIT
r
0
0
1
3504.11
5006
Command or data transmission not pos-
sible because BUSY
BIT
r
0
0
1
3504.10
5007
Command or data transmission not pos-
sible because module is faulty or ODIS is
active
BIT
r
0
0
1
3504.9
Reserve
-
BIT
r
0
0
1
3504.8
Reserve
-
BIT
r
0
0
1
3504.7
5101
Command is not permissible in this oper-
ating state
BIT
r
0
0
1
3504.6
5102
Command not possible because not at
standstill
BIT
r
0
0
1
3504.5
5104
Command not possible because range is
exceeded
BIT
r
0
0
1
3504.4
5105
Load cell parameter not plausible
BIT
r
0
0
1
3504.3
Reserve
BIT
r
0
0
1
3504.2
5107
Shifting characteristic not possible
BIT
r
0
0
1
3504.1
5199
Error in command to DI
BIT
r
0
0
1
3505.11
Summary of Contents for SIWAREX WP231
Page 16: ...Safety notes 2 1 General safety instructions SIWAREX WP231 16 Manual 07 2014 A5E31238908A 02 ...
Page 132: ......
Page 174: ...Technical data 13 3 Approvals SIWAREX WP231 174 Manual 07 2014 A5E31238908A 02 ...
Page 180: ...List of abbreviations SIWAREX WP231 180 Manual 07 2014 A5E31238908A 02 ...
Page 182: ...Index SIWAREX WP231 182 Manual 07 2014 A5E31238908A 02 ...