Operation requiring verification
12.1 Preparing for verification
SIWAREX WP231
158
Manual, 07/2014, A5E31238908A-02
12.1.4
Parameterization of the scale
The scale is parameterized prior to the verification in accordance with the envisaged use.
The scale is checked and sealed during the verification. The parameters relevant to the
verification cannot be changed after the verification. These are mainly the parameters in the
data records DR3 and DR5. These read/write parameters are identified by "rwP" (read/write
protected) in the parameter tables in Chapter 8. Certain commands cannot be used after the
verification either. These are identified in the command table by "P".
12.1.5
Adjustment and preliminary checking of the scale
During commissioning, the scale adjustment (detailed description
(Page 77)) is carried out following input of the scale parameters.
Depending on the type and the area of use, the scope of operations for checking the
measuring properties of the scale prior to the verification may be different. A preliminary
check with various calibration weights should always be carried out.
12.1.6
Calibration label
The calibration labels are edited based on the templates from the calibration set, and
subsequently printed. The fields in the calibration labels can already be provided with the
corresponding values prior to the verification. The protective foil is attached during the
verification.
Summary of Contents for SIWAREX WP231
Page 16: ...Safety notes 2 1 General safety instructions SIWAREX WP231 16 Manual 07 2014 A5E31238908A 02 ...
Page 132: ......
Page 174: ...Technical data 13 3 Approvals SIWAREX WP231 174 Manual 07 2014 A5E31238908A 02 ...
Page 180: ...List of abbreviations SIWAREX WP231 180 Manual 07 2014 A5E31238908A 02 ...
Page 182: ...Index SIWAREX WP231 182 Manual 07 2014 A5E31238908A 02 ...