Operation requiring verification
12.2 Verification
SIWAREX WP231
162
Manual, 07/2014, A5E31238908A-02
8.
Checking of the additive tare range
When using the additive tare compensation unit, you must check the complete weighing
range (up to maximum load + additive maximum tare load). To do this, you must check
up to the maximum load, and repeat this following taring. Repeat these steps until the
upper limit of the range of the additive tare compensation unit has been reached.
–
Call up the main menu in the main display (function key with the open-end wrench).
Select the "Calibration check" submenu in the main menu.
–
Select the "Scale check" submenu in the "Calibration check" submenu.
–
Check with the "Tare setting" or "Delete tare" command.
9.
Checking of the parameter disabling/calibration bridge
The calibration bridge must be inserted on the weighing module. You can check the
calibration bridge using the specific symbol
at the bottom right in the main display
(symbol displayed = calibration bridge inserted).
10.When using the second display S102 - checking of set telegram address
The telegram address (No. 01) must be set on the Siebert display S102. Checking is
carried out in accordance with the specifications in the Instruction Manual "Siebert series
S102 - digital displays for Siemens Siwarex".
Summary of Contents for SIWAREX WP231
Page 16: ...Safety notes 2 1 General safety instructions SIWAREX WP231 16 Manual 07 2014 A5E31238908A 02 ...
Page 132: ......
Page 174: ...Technical data 13 3 Approvals SIWAREX WP231 174 Manual 07 2014 A5E31238908A 02 ...
Page 180: ...List of abbreviations SIWAREX WP231 180 Manual 07 2014 A5E31238908A 02 ...
Page 182: ...Index SIWAREX WP231 182 Manual 07 2014 A5E31238908A 02 ...