Technical data
13.1 Technical specifications
SIWAREX WP231
164
Manual, 07/2014, A5E31238908A-02
Analog load cell interface connection
Table 13- 3 Technical specifications: Analog load cell interface connection
Error limit to DIN1319-1 at 20 °C +10 K
≤ 0.05% v.E.
1)
Accuracy according to EN45501 / OIML
R76
•
Class
III (and IV 1000d)
•
Resolution
(d=e)
3000d
•
Error percent-
age pi
0.4
•
Step voltage
0.5 µV/e
Accuracy delivery state
2)
typ. 0.1% v.E.
Sampling rate
100 Hz
Input signal resolution
± 4 000 000
Measuring range
± 4 mV/V
Common mode voltage range
0 V to 5 V
DMS supply
3)
4.85 V DC +2/-3 %
Short-circuit and overload protection
Yes
Connection
6-wire
Sensor voltage monitoring
≤ 0.3 V
Min. DMS input resistance
•
without Ex-i
interface
SIWAREX IS
40
Ω
•
with Ex-i inter-
face SIWAREX
IS
50
Ω
min. DMS output resistance
4 100
Ω
Temperature coefficient range
≤
± 5 ppm/K v. E.
Temperature coefficient zero point
≤
± 0.1
μV/K
Linearity error
≤
0.002 %
Measured value filtering
Low pass
Electrical isolation
500 V AC
50 Hz / 60 Hz noise suppression CMRR
> 80 dB
Input resistance
•
Signal cable
typ. 5*10
6
Ω
•
Sensor cable
typ. 60*10
6
Ω
1)
Relative accuracy! (Absolute accuracy is only reached by calibration on-site with calibration stand-
ard)
2)
Accuracy for module replacement or theoretical calibration decisive
3)
Value valid at sensor; voltage drops on cables are compensated up to 5 V
Summary of Contents for SIWAREX WP231
Page 16: ...Safety notes 2 1 General safety instructions SIWAREX WP231 16 Manual 07 2014 A5E31238908A 02 ...
Page 132: ......
Page 174: ...Technical data 13 3 Approvals SIWAREX WP231 174 Manual 07 2014 A5E31238908A 02 ...
Page 180: ...List of abbreviations SIWAREX WP231 180 Manual 07 2014 A5E31238908A 02 ...
Page 182: ...Index SIWAREX WP231 182 Manual 07 2014 A5E31238908A 02 ...