Commissioning
7.6 5 minute quick setup with SIWATOOL
SIWAREX WP231
54
Manual, 07/2014, A5E31238908A-02
7.6.5
Receive all data
Activate the "Receive all data" function in the communication menu.
All parameters can now been saved as a backup file on the hard disk. If a module is
replaced, the backup file can be downloaded to the new module within a few seconds. At the
time of input of the backup file, the scale is directly in the calibrated state again – without a
new calibration.
7.6.6
Checking the scale following calibration
If the scale is only used for company-internal purposes, a simple check is sufficient.
Perform the following steps:
1.
The scale is unloaded and shows "0 kg".
2.
Place a known reference weight on the scale.
Check the displayed value.
3.
If a second known reference weight is available, place it on the scale in addition.
Check whether the scale displays the sum of the reference weights.
4.
Remove the reference weights from the scale.
Check that the display is "0 kg" again.
Summary of Contents for SIWAREX WP231
Page 16: ...Safety notes 2 1 General safety instructions SIWAREX WP231 16 Manual 07 2014 A5E31238908A 02 ...
Page 132: ......
Page 174: ...Technical data 13 3 Approvals SIWAREX WP231 174 Manual 07 2014 A5E31238908A 02 ...
Page 180: ...List of abbreviations SIWAREX WP231 180 Manual 07 2014 A5E31238908A 02 ...
Page 182: ...Index SIWAREX WP231 182 Manual 07 2014 A5E31238908A 02 ...