Scale parameters and functions
8.7 DR 6 limit value settings
SIWAREX WP231
84
Manual, 07/2014, A5E31238908A-02
8.7
DR 6 limit value settings
The switch-on and switch-off values for the limits are configured in data record DR 6.
In legal trade operation, the data record is not write-protected.
Procedure
●
Check all parameters and modify them as required
●
Transfer the data record to the scales
Table 8- 4
Assignment of data record 6
Variable
Note
Type
Length
(bytes)
Rw
Default
Min.
Max.
Modbus
Register
Data record
number
Contains no. of data
record
USHORT
2
r
6
-
-
1234
Length
Data record length infor-
mation
USHORT
2
r
60
-
-
1235
Application
Information about which
application the DR be-
longs to
USHORT
2
r
101
-
-
1236
Version iden-
tifier
Information about the
current data record ver-
sion
USHORT
2
r
1
1
65635
1237
Basis of limits Gross / net basis of limit
values 1 and 2
0: Limit value 1 and limit
value 2 are based on the
gross value
1: Limit value 1 and limit
value 2 are based on the
net value
USHORT
2
rw
0
0
1
1238
Reserve 1
Reserve
USHORT
2
rw
0
0
-
1239
(Page 86)
Switch-on point for limit
value 1 (% of measuring
range)
FLOAT
4
rw
0
maxi-
mum
num-
ber
range
maxi-
mum
number
range
1240
(Page 86)
Time for delayed switch-
on of limit value 1 in ms
TIME
4
rw
0
0
maxi-
mum
number
range+
1242
(Page 86)
Switch-off point for limit
value 1 (% of measuring
range)
FLOAT
4
rw
0
maxi-
mum
num-
ber
range
maxi-
mum
number
range
1244
Summary of Contents for SIWAREX WP231
Page 16: ...Safety notes 2 1 General safety instructions SIWAREX WP231 16 Manual 07 2014 A5E31238908A 02 ...
Page 132: ......
Page 174: ...Technical data 13 3 Approvals SIWAREX WP231 174 Manual 07 2014 A5E31238908A 02 ...
Page 180: ...List of abbreviations SIWAREX WP231 180 Manual 07 2014 A5E31238908A 02 ...
Page 182: ...Index SIWAREX WP231 182 Manual 07 2014 A5E31238908A 02 ...