Messages
9.3 Evaluating messages with the help of SIWATOOL
SIWAREX WP231
134
Manual, 07/2014, A5E31238908A-02
9.3
Evaluating messages with the help of SIWATOOL
The electronic weighing system has a message buffer that can hold up to 80 entries. If the
number of messages in the message buffer exceeds 80, the oldest entry is overwritten. The
message buffer can be read out at any time with the help of SIWATOOL (menu item "Read
out all data records") and saved together with the scale parameters. This facilitates the
detection, analysis and correction of errors in the system.
9.4
Detecting messages with the help of FB SIWA
All messages of the SIWAREX module can be completely detected and processed in the
controller with the help of the SIWAREX WP231 function block. The messages can be
evaluated directly in a signaling system by means of a bit signaling area in the scale data
block. The message texts are stored in the signaling system. The message text is output
when a bit becomes "1".
9.5
Message list
The message list is an overview of all messages that the SIWAREX module can generate. A
message can be quickly identified by the message code (number).
Summary of Contents for SIWAREX WP231
Page 16: ...Safety notes 2 1 General safety instructions SIWAREX WP231 16 Manual 07 2014 A5E31238908A 02 ...
Page 132: ......
Page 174: ...Technical data 13 3 Approvals SIWAREX WP231 174 Manual 07 2014 A5E31238908A 02 ...
Page 180: ...List of abbreviations SIWAREX WP231 180 Manual 07 2014 A5E31238908A 02 ...
Page 182: ...Index SIWAREX WP231 182 Manual 07 2014 A5E31238908A 02 ...