Scale parameters and functions
8.3 DR 3 calibration parameters
SIWAREX WP231
72
Manual, 07/2014, A5E31238908A-02
8.3.19
Maximum negative zero setting limit (semi-automatically)
Zeroing defines the current weight of the scales as zero.
You can restrict the effect of the zeroing function by defining limits. The limitation is based
not on the current gross weight, but rather on the weight which the scales would display had
there been no zeroing (time of scale calibration).
For legal trade scales, the limitation is 4 % of the weighing range between the negative and
positive weight for zeroing.
8.3.20
Maximum positive zero setting limit (semi-automatically)
You can restrict the effect of the zeroing function by defining limits. The limitation is based
not on the current weight, but rather on the weight which the scales would display had there
been no zeroing (time of scale calibration).
For legal trade scales, the limitation is a maximum of 4 % of the weighing range between the
negative and positive weight for zeroing.