Maintenance and Calibration
•
To keep the D10/D10 Micro dry and protect the electronics from
corrosion, check/replace the desiccant bag once per year, and change
it every two years. If suspicion of any moisture inside the enclosure,
find the cause, dry it out and replace the desiccant bag. If sent to
repair after damage from moisture, please remove both main batteries
to minimize damage.
•
If the backup battery drops below 2.6 V, a warning message will be
communicated. For replacement of the backup battery, send the
instrument to Sigicom.
•
If repair is required, please contact Sigicom prior to sending the
instrument in for repair.
•
No calibration of the D10/D10 Micro is needed.