The specifications of the probe need to meet the following conditions:
1. The probe is within the validity period of calibration.
2. The ambient temperature is within 25
℃±
5
℃
.
3. The probe is correctly connected to the oscilloscope.
4. The probe and oscilloscope are in a thermally stable environment, and the
probe and oscilloscope should be warmed up for at least 20 minutes.
Warranted characteristics:
Characteristics
SAP2500D
Bandwidth (probe only)
> 2.5 GHz
Bandwidth (with scope)
2 GHz (SDS6204A)
Differential Input capacitance
1 pF
Differential Input resistance
200 k
Ω
Single-ended
Input resistance
100 k
Ω
Offset range
±8 V
Attenuation ratio (DC)
÷
10
Offset accuracy
< 3%
DC gain accuracy
< 3%
Input dynamic range
± 4 V
Maximum input voltage
(non-destructive)
20 V
Cable length
130 cm