SIGLENT
SDG1000X User Manual
133
3.7
Example
7:
Generate
a
Linear
Sweep
Waveform
Generate a sine sweep waveform whose frequency starts at 100Hz and
sweeps to a frequency of 10KHz. Use internal trigger mode, linear sweep, and
a sweep time of 2s.
Steps:
Set the sweep function.
Press Waveforms and choose the sine waveform as the sweep function.
The default setting of the source is internal.
Set the amplitude and offset.
1. Press Amplitude/HighLevel to choose Amplitude which will display in blue
color. Input ‗5‘ from the keyboard and choose the unit ‗Vpp‘ to set the
amplitude to 5Vpp.
2. Press Offset/LowLevel to choose Offset which will display in blue color.
Input ‗0‘ from the keyboard and choose the unit ‗Vdc‘ to set the offset to
0Vdc
Set the sweep time.
Press Sweep
→ Page 1/2 → Sweep Time, input ‗1‘ from the keyboard and
choose the unit
‗s‘ to set the sweep time to 1s.
Set the start frequency.
Press StartFreq, i
nput ‗100‘ from the keyboard and choose the unit ‗Hz‘ to
set the start freq to 100Hz.
Set the stop frequency.
Press StopFreq, i
nput ‗10‘ from the keyboard and choose the unit ‗kHz‘ to
set the stop freq to 10kHz.
Summary of Contents for SDG1000X Series
Page 2: ......
Page 44: ...SIGLENT 34 SDG1000X User Manual Figure 2 15 Setting the Mean...
Page 119: ...SIGLENT SDG1000X User Manual 109 Figure 2 65 Phase Deviation Interface...
Page 134: ...SIGLENT 124 SDG1000X User Manual Figure 3 1 Generate a Sine Waveform...
Page 136: ...SIGLENT 126 SDG1000X User Manual Figure 3 2 Generate a Square Waveform...
Page 148: ...SIGLENT 138 SDG1000X User Manual Figure 3 9 Generate an AM Modulation Waveform...
Page 150: ...SIGLENT 140 SDG1000X User Manual Figure 3 10 Generate a FM Modulation Waveform...
Page 152: ...SIGLENT 142 SDG1000X User Manual Figure 3 11 Generate a PM Modulation Waveform...
Page 154: ...SIGLENT 144 SDG1000X User Manual Figure 3 12 Generate a FSK Modulation Waveform...
Page 156: ...SIGLENT 146 SDG1000X User Manual Figure 3 13 Figure 3 13 Generate an ASK Modulation Waveform...
Page 158: ...SIGLENT 148 SDG1000X User Manual Figure 3 14 Generate a PSK Modulation Waveform...
Page 160: ...SIGLENT 150 SDG1000X User Manual Figure 3 15 Generate a PWM Modulation Waveform...
Page 162: ...SIGLENT 152 SDG1000X User Manual Figure 3 16 Generate a DSB AM Modulation Waveform...