SIGLENT
SDG1000X User Manual
29
To Set the Pulse Width/DutyCycle
Pulse width is defined as the time from the 50% threshold of a rising edge
amplitude to the 50% threshold of the next falling edge amplitude (as shown in
the figure below).
The pulse width setting range is limited by the ―Minimum
Pulse Width‖ and ―Pulse Period‖ setting. For detailed information, please refer
to ―SDG1000X Datasheet‖. The default value is 200μs.
Pulse duty cycle is defined as the percentage that the pulse width takes up in
the whole period. Pulse duty cycle and pulse width are correlative. Once a
parameter is changed, the other will be automatically changed.
1. Press Waveforms
→ Pulse → PulWidth, to set the pulse width.
The pulse width shown on the screen when the instrument is powered on is
the default value or the set value of last power down. If you want to set the
waveform by duty, press the PulWidth/DutyCycle key again, to switch into the
duty parameter (the current operation is displayed in inverse color).
2. Input the desired Pulse Width.
Use the numeric keyboard to input the parameter value directly, and press the
corresponding key to select the parameter unit. Or use the arrow keys to
Delay
Setting the delay for pulse waveform.
Summary of Contents for SDG1000X Series
Page 2: ......
Page 44: ...SIGLENT 34 SDG1000X User Manual Figure 2 15 Setting the Mean...
Page 119: ...SIGLENT SDG1000X User Manual 109 Figure 2 65 Phase Deviation Interface...
Page 134: ...SIGLENT 124 SDG1000X User Manual Figure 3 1 Generate a Sine Waveform...
Page 136: ...SIGLENT 126 SDG1000X User Manual Figure 3 2 Generate a Square Waveform...
Page 148: ...SIGLENT 138 SDG1000X User Manual Figure 3 9 Generate an AM Modulation Waveform...
Page 150: ...SIGLENT 140 SDG1000X User Manual Figure 3 10 Generate a FM Modulation Waveform...
Page 152: ...SIGLENT 142 SDG1000X User Manual Figure 3 11 Generate a PM Modulation Waveform...
Page 154: ...SIGLENT 144 SDG1000X User Manual Figure 3 12 Generate a FSK Modulation Waveform...
Page 156: ...SIGLENT 146 SDG1000X User Manual Figure 3 13 Figure 3 13 Generate an ASK Modulation Waveform...
Page 158: ...SIGLENT 148 SDG1000X User Manual Figure 3 14 Generate a PSK Modulation Waveform...
Page 160: ...SIGLENT 150 SDG1000X User Manual Figure 3 15 Generate a PWM Modulation Waveform...
Page 162: ...SIGLENT 152 SDG1000X User Manual Figure 3 16 Generate a DSB AM Modulation Waveform...