background image

SIGLENT

 

  140        SDG1000X User Manual

 

 

Figure 3-10 Generate a FM Modulation Waveform 

 

 

Summary of Contents for SDG1000X Series

Page 1: ...User Manual SDG1000X Series Function Arbitrary Waveform Generator UM0201X E01D 2017 SIGLENT TECHNOLOGIES CO LTD...

Page 2: ......

Page 3: ...SIGLENT SDG1000X User Manual I Declaration Copyright SIGLENT TECHNOLOGIES CO LTD All rights reserved Without permission contents in this manual are not allowed to be copied extracted or translated...

Page 4: ...the instrument is grounded correctly before connecting its input or output terminals Connect the signal wire correctly The potential of the signal wire ground is equal to the earth therefore do not co...

Page 5: ...ry or hazard that may immediately happen WARNING Indicates an injury or hazard that may not immediately happen CAUTION Indicates that a potential damage to the instrument or other property might occur...

Page 6: ...th these advantages the SDG1000X can provide users with a variety of high fidelity and low jitter signals and can meet the growing requirements of complex and extensive applications Key Features Dual...

Page 7: ...SIGLENT SDG1000X User Manual V High precision Frequency Counter 196 kinds of built in arbitrary waveforms Standard interfaces USB Host USB Device USBTMC LAN VXI 11 Optional interface GPIB 4 3 display...

Page 8: ...6 2 Front Panel Operations 17 2 1 To Set Sine Waveform 18 2 2 To Set Square Waveform 23 2 3 To Set Ramp Waveform 26 2 4 To Set Pulse Waveform 28 2 5 To Set Noise Waveform 32 2 6 To Set DC Waveform 35...

Page 9: ...ulse Waveform 129 3 5 Example 5 Generate a Noise 131 3 6 Example 6 Generate a DC Waveform 132 3 7 Example 7 Generate a Linear Sweep Waveform 133 3 8 Example 8 Generate a Burst Waveform 135 3 9 Example...

Page 10: ...SIGLENT VIII SDG1000X User Manual 5 1 Maintenance summary 155 5 2 Contact SIGLENT 156 6 Appendix 157 Appendix A Accessories 157 Appendix B Daily Maintain and Cleaning 158...

Page 11: ...anual 1 1 Quick Start This chapter covers the following topics Handle Adjustment The Front Rear Panel To Select a Waveform To Set Modulation Sweep Burst To Turn On Off Output To Use Numeric Input To U...

Page 12: ...al 1 1 Handle Adjustment To adjust the handle position of the SDG1000X please grip the handle by the sides and pull it outward Then make the handle rotate to the desired position Figure 1 1 Viewing Po...

Page 13: ...ns of the front rear panel Front Panel SDG1000X has a clear and simple front panel which includes a 4 3 inch screen menu softkeys numeric keyboard knob function keys arrow keys and channel control are...

Page 14: ...des multiple interfaces including Counter 10MHz In Out Aux In Out LAN USB Device Earth Terminal and AC Power Supply Input Figure 1 3 Rear Panel of SDG1000X Aux In Out 10 MHz Clock Input Output Earth T...

Page 15: ...e function selected Figure 1 4 User Interface 1 Waveform Display Area Displays the currently selected waveform of each channel 2 Channel Status Bar Indicates the selected status and output configurati...

Page 16: ...he instrument s output impedance of 50 but rather is used to maintain amplitude accuracy into different load values Output Channel output state After pressing corresponding channel output control port...

Page 17: ...ual 7 1 4 shows the parameters of AM modulation 8 Modulation Parameters Area Shows the parameters of the current modulation function After selecting the corresponding menu use number keys or knob to c...

Page 18: ...with frequencies from 1 Hz to 60MHz By setting Frequency Period Amplitude High level Offset Low level and Phase a sine waveform with different parameters can be generated Figure 1 6 Sine Display Inte...

Page 19: ...y Period Amplitude High level Offset Low level Phase and Symmetry a ramp waveform with different parameters can be generated Figure 1 8 Ramp Display Interface 4 Press Waveforms key and then press Puls...

Page 20: ...ress Noise softkey The generator can generate noise with a 60MHz bandwidth By setting Stdev and Mean noise with different parameters can be generated Figure 1 10 Noise Display Interface 6 Press Wavefo...

Page 21: ...t press the Arb softkey The generator can generate repeatable arbitrary waveforms with 16K points and frequencies up to 6MHz By setting Frequency Period Amplitude High level Offset Low level and Phase...

Page 22: ...fying the parameters such as Type Source AM Depth AM Freq Shape etc The SDG1000X can modulate waveforms using AM FM PM ASK FSK PSK PWM and DSB AM etc Pulse waveforms can only be modulated using PWM No...

Page 23: ...lay Interface 3 Press Burst the Burst function will be enabled Burst signals for sine square ramp pulse or arbitrary waveforms may be generated Start Phase ranges from 0 to 360 and Burst Period ranges...

Page 24: ...he output of the two channels Choose a channel and press the corresponding Output key the key backlight will be lighted and the output will be enabled Press the Output key again the key backlight will...

Page 25: ...the digital input selection Figure 1 18 Front Panel Digital Input 1 The numeric keyboard is used to enter the parameter s value 2 The knob is used to increase clockwise or decrease counterclockwise t...

Page 26: ...et the parameters of basic waveforms directly 2 The Utility key is used to set the auxiliary system function such as output configurations interface setting system setting information performing the i...

Page 27: ...apter 1 once more This chapter covers the following topics To Set Sine To Set Square To Set Ramp To Set Pulse To Set Noise To Set DC To Set Arbitrary To Set Harmonic Function To Set Modulation Functio...

Page 28: ...menu select Frequency The frequency parameter area is highlighted in the parameter display window and users can set the frequency value here Figure 2 1 Sine Parameters Display Interface Table2 1 Menu...

Page 29: ...ument is powered on is the default value or the set value of last power down If Period rather than Frequency is the desired parameter press Frequency Period again to enter the Period mode The current...

Page 30: ...amplitude The amplitude shown on the screen when the instrument is powered on is the default value or the set value of last power down If setting the waveform s high level is desired press the Amplit...

Page 31: ...lue or the set value of last power down If you want to set the waveform by low level press the Offset LowLevel key again to switch into the low level parameter the current operation is displayed in in...

Page 32: ...2 Input the desired phase Use the numeric keyboard to input the parameter value directly and press the corresponding key to select the parameter unit Or use the arrow keys to select the digit to edit...

Page 33: ...he parameter display window and users can set the duty cycle value here Figure 2 6 Square Parameters Display Interface Table2 2 Menu Explanations of Square Waveform Function Menu Settings Explanation...

Page 34: ...lt value is 50 1 Press Waveforms Square DutyCycle to set the duty cycle The duty cycle shown on the screen when the instrument is powered on is the default value or the set value of last power down 2...

Page 35: ...SIGLENT SDG1000X User Manual 25 Figure 2 7 Setting the Duty Cycle Note The methods of setting other parameters of square signal are similar to sine waveform function...

Page 36: ...lighted in the parameter display window and users can set the symmetry value here Figure 2 8 Ramp Parameters Display Interface Table2 3 Menu Explanations of Ramp Waveform Function Menu Settings Explan...

Page 37: ...the default value or the set value of last power down 2 Input the desired Symmetry Use the numeric keyboard to input the parameter value directly and press the corresponding key to select the paramet...

Page 38: ...alue here Figure 2 10 Pulse Parameters Display Interface Table 2 4 Menu Explanations of Pulse Waveform Function Menu Settings Explanation Frequency Period Set the signal frequency or period The curren...

Page 39: ...le period Pulse duty cycle and pulse width are correlative Once a parameter is changed the other will be automatically changed 1 Press Waveforms Pulse PulWidth to set the pulse width The pulse width s...

Page 40: ...dge time is limited by the currently specified pulse width limit Users can set rise edge and fall edge independently 1 Press Waveforms Pulse Rise to set the rise edge The rise edge shown on the screen...

Page 41: ...e digit to edit and then use the knob to change its value The generator will change the waveform immediately Figure 2 12 Setting the Rise Edge Note The methods of setting other parameters of pulse sig...

Page 42: ...ea is highlighted in the parameter display window and users can set the stdev value here Noise is non periodic signal which has no frequency or period Figure 2 13 Noise Parameters Display Interface Ta...

Page 43: ...its value Figure 2 14 Setting the Stdev To Set the Mean 1 Press Waveforms Noise Mean to set the mean The mean shown on the screen when the instrument is powered on is the default value or the set val...

Page 44: ...SIGLENT 34 SDG1000X User Manual Figure 2 15 Setting the Mean...

Page 45: ...Press Waveforms Page 1 2 DC to enter the following interface Please note that there is a DC offset parameter at the middle of the screen Figure 2 16 DC Setting Interface Note The method of setting of...

Page 46: ...nd phase Figure 2 17 Arb Parameters Display Interface DDS Table 2 6 Menu Explanations of Arb Waveform Page 1 2 Function menu Settings Explanations Frequency Period Set the signal frequency or period T...

Page 47: ...lowing interface as shown in Figure 2 18 Figure 2 18 Built in Arbitrary Waveforms Press Common Math Engine Window Trigo or other menus to switch to the desired category the selected category in the me...

Page 48: ...3 X3 function Airy Airy function Besselj Bessel I function Bessely Bessel II function Dirichlet Dirichlet function Erf Error function Erfc Complementary error function ErfcInv Inverted complementary e...

Page 49: ...locity curve of explosive oscillation DampedOsc Time displacement curve of damped oscillation SwingOsc Kinetic energy time curve of swing oscillation Discharge Discharge curve of NI MH battery Pahcur...

Page 50: ...attopWin Flat top weighted window ParzenWin ParzenWin window TaylorWin TaylorWin window TukeyWin TukeyWin tapered cosine window Trigo Tan Tangent Cot Cotangent Sec Secant Csc Cosecant Asin Arc sine Ac...

Page 51: ...y SquareDuty26 Square waveform with 26 duty SquareDuty28 Square waveform with 28 duty SquareDuty30 Square waveform with 30 duty SquareDuty32 Square waveform with 32 duty SquareDuty34 Square waveform w...

Page 52: ...e waveform with 88 duty SquareDuty90 Square waveform with 90 duty SquareDuty92 Square waveform with 92 duty SquareDuty94 Square waveform with 94 duty SquareDuty96 Square waveform with 96 duty SquareDu...

Page 53: ...Sectional sine AM signal FM Sectional sine FM signal PFM Sectional pulse FM signal PM Sectional sine PM signal l PWM Sectional PWM signal Filter Butterworth Butterworth filter Chebyshev1 Chebyshev1 f...

Page 54: ...SIGLENT 44 SDG1000X User Manual Figure 2 19 Stored Waveform Display Interface Rotate the knob to choose the desired waveform Then select Recall or press the knob to recall the corresponding waveform...

Page 55: ...l waveform frequency A1 is fundamental waveform amplitude and 1 is fundamental waveform phase The frequencies of the other components called harmonics are all integral multiples of the fundamental wav...

Page 56: ...onics 2 Press Odd the instrument will output fundamental waveform and odd harmonics 3 Press All the instrument will output fundamental waveform and all the user defined orders of harmonics To Set the...

Page 57: ...pop up menu The units available are Vpp mVpp and dBc To Select the Harmonic Phase After entering the harmonic setting menu press Harmonic Phase to set the harmonic phase of each order 1 Press Order t...

Page 58: ...e source internal external modulating frequency frequency deviation modulating waveform and carrier In PM users can set the source internal external phase deviation modulating frequency modulating wav...

Page 59: ...he parameters of AM modulation are shown in Figure 2 21 Figure 2 21 Setting Interface of AM Modulation Table 2 9 Menu Explanations of the AM Parameters Function Menu Settings Explanation Type AM Ampli...

Page 60: ...orm selected of the current channel Note Noise can be used as modulating waveform but cannot be used as the carrier 2 External Source When external AM modulation source is selected the generator accep...

Page 61: ...epth to set the parameter In the 0 modulation the output amplitude is the half of the carrier s amplitude In the 120 modulation the output amplitude is the same with the carrier s amplitude For an ext...

Page 62: ...f DSB AM Modulation Table 2 10 Menu Explanations of the DSB AM Parameters Function Menu Settings Explanation Type DSB AM DSB Amplitude modulation Source Internal The source is internal External The so...

Page 63: ...SIGLENT SDG1000X User Manual 53 The methods of setting the parameters of DSB AM are similar to AM...

Page 64: ...parameters of FM modulation are shown in Figure 2 23 Figure 2 23 Setting Interface of FM Modulation Table 2 11 Menu Explanations of the FM Parameters Function Menu Settings Explanation Type FM Freque...

Page 65: ...ight the parameter and then use the numeric keyboard or arrow keys and knob to input the desired value The deviation should be equal to or less than the carrier frequency The sum of the deviation and...

Page 66: ...parameters of PM modulation are shown in Figure 2 24 Figure 2 24 Setting Interface of PM Modulation Table 2 12 Menu Explanations of the PM Parameters Function Menu Settings Explanation Type PM Phase m...

Page 67: ...hase Dev to highlight the parameter and then use the numeric keyboard or arrow keys and knob to input the desired value Use the numeric keyboard or arrow keys and knob to input the desired value The r...

Page 68: ...ace of FSK Modulation Table 2 13 Menu Explanations of the FSK Parameters Function Menu Settings Explanation Type FSK Frequency shift keying modulation Source Internal The source is internal External T...

Page 69: ...hidden To Set Hop Frequency The range of the hop frequency depends on the carrier frequency currently selected Press Hop Freq to highlight the parameter and then use the numeric keyboard or arrow keys...

Page 70: ...planations of the ASK Parameters Function Menu Settings Explanation Type ASK Amplitude shift keying modulation Source Internal The source is internal External The source is external Use the Aux In Out...

Page 71: ...in Figure 2 27 Figure 2 27 Setting Interface of PSK Modulation Table 2 15 Menu Explanations of the PSK Parameters Function Menu Settings Explanation Type PSK Phase shift keying modulation Source Inter...

Page 72: ...62 SDG1000X User Manual Note The methods of setting the parameters of PSK are similar to AM In addition the external modulating signal of PSK must be Square which complies with the CMOS level specifi...

Page 73: ...own in Figure 2 28 Figure 2 28 Setting Interface of PWM Modulation Table 2 16 Menu Explanations of the PWM Parameters Function Menu Settings Explanation Type PWM Pulse width modulation The carrier is...

Page 74: ...numeric keyboard or arrow keys and knob to input the desired value as shown in the Figure 2 29 Figure 2 29 Width Deviation Setting Interface The width deviation cannot exceed the current pulse width T...

Page 75: ...ed the current pulse duty cycle The duty deviation is limited by the minimum duty cycle and current edge time setting Duty deviation and width deviation are correlative Once a parameter is changed the...

Page 76: ...2 17 Menu Explanations of Sweep Page 1 2 Function Menu Settings Explanation Sweep time Set the time span of the sweep in which the frequency changes from the start frequency to stop frequency Start Fr...

Page 77: ...key again to switch between the two sweep range modes Start Frequency and Stop Frequency Start Frequency and Stop Frequency are the lower and upper limits of the frequency for sweep Start Frequency S...

Page 78: ...ency Stop Frequency 2 Frequency Span Stop Frequency Start Frequency Sweep Type SDG1000X provides Linear and Log sweep profiles and the default is Linear Linear Sweep In linear sweep the output frequen...

Page 79: ...ode Figure 2 34 Log Sweep Interface Sweep Trigger Source The sweep trigger source can be internal external or manual The generator will generate a sweep output when a trigger signal is received and th...

Page 80: ...eives a CMOS pulse with specified polarity To set the CMOS pulse polarity choose Edge to select Up or Down 3 Manual Trigger Choose Source Manual a sweep will be generated from the corresponding channe...

Page 81: ...ong burst type trigger source and carrier Burst Type Trigger Source Carrier N Cycle Internal External Manual Sine Square Ramp Pulse Arbitrary Infinite External Manual Sine Square Ramp Pulse Arbitrary...

Page 82: ...e Cycles Infinite Set the number of the bursts in N Cycle Set the number of the bursts in N Cycle to be infinite Start Phase Set the start phase of the burst Burst Period Set the burst period Source I...

Page 83: ...Enable trigger out Page 2 2 Return to the previous page Infinite In infinite mode the cycle number of the waveform is set as an infinite value The generator outputs a continuous waveform after receiv...

Page 84: ...level When the gated signal is true the generator outputs a continuous waveform When the gated signal is false the generator first completes the output of the current period and then stops Waveforms...

Page 85: ...he default setting is 0 For an Arbitrary Waveform 0 is the first waveform point Burst Period Burst Period is only available when the trigger source is internal It is defined as the time from the start...

Page 86: ...ger is selected Choose Trig Out as Up or Down the Aux In Out connector at the rear panel will output a trigger signal with specified edge 2 External Trigger Choose Source External the generator accept...

Page 87: ...ing interface Figure 2 39 Store Recall Interface Page 1 2 Table 2 23 Menu Explanations of Save and Recall Function Menu Settings Explanation File Type State The setting of the generator Data Arbitrary...

Page 88: ...ary waveform files to C Disk 2 USB Device 0 There is a USB Host interface located on the left side of the front panel which permits users to store recall waveforms or update the firmware version by U...

Page 89: ...re used the name may be displayed in the store and recall interface abnormally Browse Use the knob to shift between the directories to choose Local C or USB Device 0 Choose Browse press the knob to op...

Page 90: ...Data File The SDG1000X can recall the data files in csv or dat format from the external memory and transfer them into bin format then store them in the internal memory When it is done the generator wi...

Page 91: ...c waveform parameters modulation parameters and other utility settings used To save the instrument state the procedures are given as followed 1 Choose the file type to store Press Store Recall File Ty...

Page 92: ...Down menus Then choose Select to display the character selected in the filename area Delete the character Use the left and right arrow keys to move the cursor in the file name Then choose Delete to de...

Page 93: ...the file Choose Delete the generator will display prompt message Delete the file Then press Accept the generator will delete the currently selected file To Copy and Paste File SDG1000X supports the in...

Page 94: ...SIGLENT 84 SDG1000X User Manual 3 Paste the file Rotate the knob to select Local C and press the knob to open its directory Then press Paste...

Page 95: ...y menu as shown in Figure 2 43 and Figure 2 44 Figure 2 43 Utility Setup Interface Page 1 2 Table 2 26 Menu Explanations of Utility Page1 2 Function Menu Settings Explanation System Set the system con...

Page 96: ...2 Function Menu Settings Explanation Interface Set the parameters of remote interfaces Sync Set the sync output CLKSource Internal Choose the system clock source External Mode Choose Phase locked or i...

Page 97: ...System Setup Page 1 2 Function Menu Settings Explanation Number format Set the number format Language English Set the language Chinese PowerOn Default All the settings return to default when power on...

Page 98: ...able the screen saver 5min 15min 30min 1hour 2hour 5hour Off Disable the screen saver System Info View the system information Firmware Update Update the firmware by the U disk Help View the help infor...

Page 99: ...Close the Separator Space Use Space as the separator Accept Save the current settings and return to the System menu According to the different choices of the decimal point and the separator the forma...

Page 100: ...the SDG1000X s setting when the generator is powered on Two choices are available the default setting and the last settings set when the unit was last powered down Once selected the setting will be a...

Page 101: ...ettings of the system are as followed Table 2 31 Factory Default Setting Output Default Function Sine Wave Frequency 1kHz Amplitude Offset 4Vpp 0Vdc Phase 0 Load High Z Modulation Default Carrier 1kHz...

Page 102: ...nd the default is On 6 Screen Saver Enable or disable screen saver Press Utility System Page 1 2 ScrnSvr to select On or Off and the default is Off Screen saver will be on if no action is taken within...

Page 103: ...or 2 Press Utility Page 1 2 Firmware Update Or press Store Recall directly 3 Select the firmware file ADS and choose Recall to update the system software 4 After the updating is finished the generator...

Page 104: ...the help information at any time when operating the instrument Press Utility System Page 1 2 Help to enter the following interface Figure 2 51 Help Menu Table 2 32 Help Menu Explanations Function Men...

Page 105: ...the following interface Figure 2 52 Test Cal function Menu Table 2 33 Menu Explanations of Test Cal Setting Function Menu Settings Explain SelfTest Perform a system self test Cancel Return to the Uti...

Page 106: ...test program LEDTest Run key indicator lights test program BoardTest Run hardware circuit self test program Cancel Return to the Test Cal menu 1 ScrTest Select ScrTest to enter the screen test interfa...

Page 107: ...e circle between two arrows represents the knob Test all keys and knob and also verify that all the backlight keys illuminate correctly Figure 2 55 Key Test Interface The corresponding area of tested...

Page 108: ...ey to continue press 8 Key to exit is displayed Press the 7 key continuously for testing and when a key is lighted the corresponding area on the screen will display in blue color Figure 2 56 LED Test...

Page 109: ...ace Table 2 35 Menu Explanations of Frequency Counter Function Menu Settings Explanation State Off Open the counter On Close the counter Frequency Measured frequency Period Measured period PWidth Meas...

Page 110: ...to default Accept Save the current settings and return to the previous menu 1 To Select the Parameters to be measured The frequency counter on the SDG1000X can measure parameters including frequency...

Page 111: ...The default is AC 5 HFR High Frequency Rejection can be used to filter out the high frequency components of a measured signal and improve the measurement accuracy in low frequency signal measurement...

Page 112: ...ltage will not be the same as the output voltage setting on the generator This function is used to match the displayed voltage with the expected one This setting does not actually change the output im...

Page 113: ...t inverted when the waveform is inverted EqPhase Press Utility Output Setup EqPhase to align the phases of CH1 and CH2 Choosing the menu will re configure two channels and enable the generator to outp...

Page 114: ...Explanation CH1 Switch CH1 Output the waveform of CH1 CH1 CH2 Output the waveform of CH1 CH2 CH2 Switch CH2 Output the waveform of CH2 CH1 CH2 Output the waveform of CH1 CH2 Return Save the current o...

Page 115: ...one channel to the other one Press Utility CH Copy Coupling Channel Copy to enter the following interface Figure 2 62 Channel Copy Interface Table 2 38 Menu Explanations of Channel Copy Function Menu...

Page 116: ...ng is enabled CH1 and CH2 can be modified simultaneously When the frequency amplitude or phase of one channel as the reference is changed the corresponding parameter of the other channel will be chang...

Page 117: ...itude deviation between CH1 and CH2 The resulting signal is represented by AmplCH2 AmplCH1 AmplDev Ratio the amplitude ratio of CH1 and CH2 The resulting signal is represented by AmplCH2 AmplCH1 AmplR...

Page 118: ...r states of CH2 will be adjusted to the same values or states automatically At this point the dual channels can output the same signal Choose Utility CH Copy Coupling Track to enable or disable the tr...

Page 119: ...SIGLENT SDG1000X User Manual 109 Figure 2 65 Phase Deviation Interface...

Page 120: ...rameters or GPIB address Figure 2 66 Interface Settings Table 2 39 Menu Explanations of Interface Function Menu Settings Explanation GPIB Set the GPIB address LAN State On Turn on LAN Off Turn off LAN...

Page 121: ...uggested to do as the following steps 1 Connect the device Connect the USB Device interface at the rear panel of SDG1000X with the PC via a USB cable 2 Install the USB driver NI Visa is recommended 3...

Page 122: ...terminal of the PC 2 Install the driver of GPIB card Please install the driver for the GPIB card which has been connected to your PC 3 Set the GPIB address Choose Utility Page 1 2 Interface GPIB to e...

Page 123: ...he device Connect the generator to your PC or the LAN of your PC using a network cable 2 Configure network parameters Choose Utility Page 1 2 Interface LAN State to turn on LAN Then choose LAN Setup t...

Page 124: ...d on at the next time 3 To Set Gateway The format of gateway is nnn nnn nnn nnn and each nnn ranges from 0 to 255 It is recommended to acquire an available gateway from your network administrator Pres...

Page 125: ...DHCP to select On or Off to turn DHCP mode on or off The default is Off 3 Communicate with PC remotely Open Measurement Automation Explorer of NI After adding the LAN device VISA TCP IP Resource succe...

Page 126: ...ut Interface Table 2 40 Menu Explanations of Sync Output Function Menu Settings Explanation State Off Close the sync output On Open the sync output Channel CH1 Set the sync signal of CH1 CH2 Set the s...

Page 127: ...ulse with about 50ns pulse width For AM FM PM and PWM the frequency of the sync signal is the modulating frequency For ASK FSK and PSK the frequency of the sync signal is the key frequency 2 When exte...

Page 128: ...ear panel If not the prompt message No external clock source would be displayed and the clock source would be switched to Internal Sync methods for two or more instruments Synchronization between two...

Page 129: ...enter the mode setup Interface as shown in Figure 2 70 Mode Setup Interface Figure 2 70 Mode Setup Interface Phase locked Mode When changing the frequency the DDSs of both channels reset and the phas...

Page 130: ...the frequency neither channels DDS resets and the phase deviation between CH1 and CH2 changes at random When the independent mode is enabled the phase parameter cannot be modified and the menu Phase...

Page 131: ...H2 will take effect once any of the following conditions is met When overvoltage protection occurs a message will be displayed and the output is disabled The absolute value of input voltage is higher...

Page 132: ...ate a Ramp Waveform Example 4 Generate a Pulse Waveform Example 5 Generate a Noise Example 6 Generate a DC Waveform Example 7 Generate a Linear Sweep Waveform Example 8 Generate a Burst Waveform Examp...

Page 133: ...and choose the unit MHz The frequency is set to 1MHz Set the Amplitude 1 Press Amplitude HighLevel to choose Amplitude which will display in blue color 2 Input 5 from the keyboard and choose the unit...

Page 134: ...SIGLENT 124 SDG1000X User Manual Figure 3 1 Generate a Sine Waveform...

Page 135: ...ess Amplitude HighLevel to choose Amplitude which will display in blue color 2 Input 2 from the keyboard and choose the unit Vpp The amplitude is set to 2Vpp Set the Offset 1 Press Offset LowLevel to...

Page 136: ...SIGLENT 126 SDG1000X User Manual Figure 3 2 Generate a Square Waveform...

Page 137: ...Amplitude which will display in blue color 2 Input 100 from the keyboard and choose the unit mVpp The amplitude is set to 100mVpp Set the Offset 1 Press Offset LowLevel to choose Offset which will di...

Page 138: ...SIGLENT 128 SDG1000X User Manual When the period amplitude offset phase and symmetry are set the waveform generated is shown in Figure 3 3 Figure 3 3 Generate a Ramp Waveform...

Page 139: ...1 Press Amplitude HighLevel and choose the HighLevel which will display in blue color 2 Input 5 from the keyboard and choose the unit V The high level is set to 5V Set the LowLevel 1 Press Offset Low...

Page 140: ...l 2 Input 20 from the keyboard and choose the unit ns The delay is set to 20ns When the frequency high level low level pulse width and delay are set the waveform generated is shown in Figure 3 4 Figur...

Page 141: ...e Stdev which will display in blue color 2 Input 0 5 from the keyboard and choose the unit V The stdev is set to 0 5 V Set the Mean 1 Press Mean to choose Mean which will display in blue color 2 Input...

Page 142: ...aveform 1 Press Waveforms Page 1 2 DC to choose the DC waveform Set the Offset 1 Press Offset and choose Offset which will display in blue color 2 Input 3 from the keyboard and choose the unit Vdc The...

Page 143: ...oose Amplitude which will display in blue color Input 5 from the keyboard and choose the unit Vpp to set the amplitude to 5Vpp 2 Press Offset LowLevel to choose Offset which will display in blue color...

Page 144: ...SDG1000X User Manual Set the sweep profiles Press Type and choose Linear When all parameters above are set the linear sweep waveform generated is shown in Figure 3 7 Figure 3 7 Generate a Linear Swee...

Page 145: ...ncy to 10kHz 2 Press Amplitude HighLevel to choose Amplitude which will display in blue color Input 4 from the keyboard and choose the unit Vpp to set the amplitude to 4Vpp 3 Press Offset LowLevel to...

Page 146: ...e the unit Cycle to set the burst cycle count to 5 Set the delay Press Page 1 2 to choose Delay and input 100 from the keyboard and choose the unit s to set the delay to 100 s When all parameters abov...

Page 147: ...choose Amplitude which will display in blue color Input 1 from the keyboard and choose the unit Vpp to set the amplitude to 1Vpp 4 Press Offset LowLevel and choose Offset which will display in blue co...

Page 148: ...SIGLENT 138 SDG1000X User Manual Figure 3 9 Generate an AM Modulation Waveform...

Page 149: ...and choose Amplitude which will display in blue color Input 1 from the keyboard and choose the unit Vpp to set the amplitude to 1Vpp 4 Press Offset LowLevel and choose Offset which will display in blu...

Page 150: ...SIGLENT 140 SDG1000X User Manual Figure 3 10 Generate a FM Modulation Waveform...

Page 151: ...choose Amplitude which will display in blue color Input 5 from the keyboard and choose the unit Vpp to set the amplitude to 5Vpp 4 Press Offset LowLevel and choose Offset which will display in blue co...

Page 152: ...SIGLENT 142 SDG1000X User Manual Figure 3 11 Generate a PM Modulation Waveform...

Page 153: ...ghLevel and choose Amplitude which will display in blue color Input 5 from the keyboard and choose the unit Vpp to set the amplitude to 5Vpp 4 Press Offset LowLevel and choose Offset which will displa...

Page 154: ...SIGLENT 144 SDG1000X User Manual Figure 3 12 Generate a FSK Modulation Waveform...

Page 155: ...he frequency to 5kHz 3 Press Amplitude HighLevel and choose Amplitude which will display in blue color Input 5 from the keyboard and choose the unit Vpp to set the amplitude to 5Vpp 4 Press Offset Low...

Page 156: ...SIGLENT 146 SDG1000X User Manual Figure 3 13 Figure 3 13 Generate an ASK Modulation Waveform...

Page 157: ...ress Amplitude HighLevel and choose Amplitude which will display in blue color Input 5 from the keyboard and choose the unit Vpp to set the amplitude to 5Vpp 4 Press Offset LowLevel and choose Offset...

Page 158: ...SIGLENT 148 SDG1000X User Manual Figure 3 14 Generate a PSK Modulation Waveform...

Page 159: ...board and choose the unit Vpp to set the amplitude to 5Vpp 4 Press Offset LowLevel and choose Offset which will display in blue color Input 0 from the keyboard and choose the unit Vdc to set the offse...

Page 160: ...SIGLENT 150 SDG1000X User Manual Figure 3 15 Generate a PWM Modulation Waveform...

Page 161: ...frequency to 2kHz 3 Press Amplitude HighLevel and choose Amplitude which will display in blue color Input 4 from the keyboard and choose the unit Vpp to set the amplitude to 4Vpp 4 Press Offset LowLev...

Page 162: ...SIGLENT 152 SDG1000X User Manual Figure 3 16 Generate a DSB AM Modulation Waveform...

Page 163: ...hanically and electrically 2 Inspect the entire instrument In case there is any mechanical damage or defect or the instrument does not operate properly or fails in the performance tests notify the SIG...

Page 164: ...n 3 After the inspections above restart the generator 4 If the generator still doesn t work after checking please contact SIGLENT 2 If there is no waveform output after setting the parameters please d...

Page 165: ...SIGLENT will provide repair or replace the unit as described in the complete warranty statement To arrange for service or obtain a copy of the complete warranty statement please contact your nearest...

Page 166: ...al 5 2 Contact SIGLENT SIGLENT TECHNOLOGIES CO LTD Address 3 F NO 4 building Antongda Industrial Zone 3rd Liuxian Road 68th District Baoan District Shenzhen P R China Tel 400 878 0807 E mail sales sig...

Page 167: ...ries Function Arbitrary Waveform Generator Accessories Standard Accessories Quick Start Guide A Calibration Report A Power Line that fits the standard of the destination country A USB Cable Optional A...

Page 168: ...t from all power sources and clean it with a mild detergent and water Make sure the instrument is completely dry before reconnecting it to a power source To clean the exterior surface perform the foll...