SDG6000X User Manual 145
2.8.10 Overvoltage Protection
Press Utility
→ Page 1/3 → OverVoltage Protection to turn on or off the function,
as shown in the following figure.
Figure 2-104 Overvoltage Protection Interface
If the state is set to ON, overvoltage protection of CH1 and CH2 will take effect
once any of the following conditions is met:
The absolute value of input voltage is higher than 11V±0.5 V when the
amplitude of the generator is higher than or equal-to 3.2 Vpp or the DC
offset is greater-than or equal-to |2 VDC|.
The absolute value of input voltage is higher than 4 V ± 0.5 V when the
amplitude of the generator is lower than 3.2 Vpp and the DC offset is lower
than |2 VDC|.
When overvoltage protection occurs, a message will be displayed and the
output is disabled.
Summary of Contents for SDG6000X Series
Page 2: ......
Page 45: ...SDG6000X User Manual 35 Figure 2 16 Setting the Bandwidth...
Page 142: ...132 SDG6000X User Manual Figure 2 95 Phase Deviation Interface...
Page 160: ...150 SDG6000X User Manual Figure 3 1 Generate a Sine Waveform...
Page 162: ...152 SDG6000X User Manual Figure 3 2 Generate a Square Waveform...
Page 175: ...SDG6000X User Manual 165 Figure 3 9 Generate an AM Modulation Waveform...
Page 177: ...SDG6000X User Manual 167 Figure 3 10 Generate a FM Modulation Waveform...
Page 179: ...SDG6000X User Manual 169 Figure 3 11 Generate a PM Modulation Waveform...
Page 181: ...SDG6000X User Manual 171 Figure 3 12 Generate a FSK Modulation Waveform...
Page 183: ...SDG6000X User Manual 173 Figure 3 13 Generate an ASK Modulation Waveform...
Page 185: ...SDG6000X User Manual 175 Figure 3 14 Generate a PSK Modulation Waveform...
Page 187: ...SDG6000X User Manual 177 Figure 3 15 Generate a PWM Modulation Waveform...
Page 189: ...SDG6000X User Manual 179 Figure 3 16 Generate a DSB AM Modulation Waveform...