SDG6000X User Manual 59
waveform. Rotate the knob to select the .arb file and then push down the knob
or press the Recall key to recall the IQ data. The Browse key is used to enter
the subdirectory, and the Delete key is used to delete the selected file.
Note:
When the EasyIQ downloads a waveform from PC to the instrument,
the instrument will store the waveform as a .arb file. The file can also be
copied to the instrument by a USB flash disk.
I/Q Adjustment
Choose Parameter
→I/Q Adjustment to enter the adjustment interface, as
show in Figure 2-35. The adjust parameters include Gain Balance, I/Q Offset
and Angle Adjustment. Detailed description is show in the table below:
Table 2-11
IQ Adjustment Parameters
Parameter
Explanation
Gain Balance
Amplitude gain balance, adjusting the difference between I and Q
amplitude, range : -4 dB~4 dB
I/Q Offset
I or Q DC offset, range: -0.25 Vdc~0.25 Vdc
Angle Adjustment
Not supported for now
Figure 2-35 Setting I/Q Adjustment
Summary of Contents for SDG6000X Series
Page 2: ......
Page 45: ...SDG6000X User Manual 35 Figure 2 16 Setting the Bandwidth...
Page 142: ...132 SDG6000X User Manual Figure 2 95 Phase Deviation Interface...
Page 160: ...150 SDG6000X User Manual Figure 3 1 Generate a Sine Waveform...
Page 162: ...152 SDG6000X User Manual Figure 3 2 Generate a Square Waveform...
Page 175: ...SDG6000X User Manual 165 Figure 3 9 Generate an AM Modulation Waveform...
Page 177: ...SDG6000X User Manual 167 Figure 3 10 Generate a FM Modulation Waveform...
Page 179: ...SDG6000X User Manual 169 Figure 3 11 Generate a PM Modulation Waveform...
Page 181: ...SDG6000X User Manual 171 Figure 3 12 Generate a FSK Modulation Waveform...
Page 183: ...SDG6000X User Manual 173 Figure 3 13 Generate an ASK Modulation Waveform...
Page 185: ...SDG6000X User Manual 175 Figure 3 14 Generate a PSK Modulation Waveform...
Page 187: ...SDG6000X User Manual 177 Figure 3 15 Generate a PWM Modulation Waveform...
Page 189: ...SDG6000X User Manual 179 Figure 3 16 Generate a DSB AM Modulation Waveform...