SDG6000X User Manual 89
frequency for sweep.
Start Frequency ≤ Stop Frequency.
Choose Direction
→ Up, the generator will sweep from Start frequency to
Stop frequency.
Choose Direction
→ Down, the generator will sweep from Stop frequency
to Start frequency.
Choose Direction
→ Up_Down, the generator will sweep from Start
frequency to Stop frequency, then from Stop frequency to Start frequency.
Center Frequency and Frequency Span
Center Frequency = (|Start Fre Stop Frequency|)/2
Frequency Span = Stop Frequency
– Start Frequency
Sweep Type
SDG6000X provides
“Linear” and “Log” sweep profiles and the default is
“Linear”.
Linear Sweep
In linear sweep, the output frequency of the instrument varies linearly in the
way of “a number of Hertz per second”. Choose Sweep → Page 1/2 → Type
→ Linear, there is a straight line displayed on the waveform on the screen,
indicating that the output frequency varies linearly.
Summary of Contents for SDG6000X Series
Page 2: ......
Page 45: ...SDG6000X User Manual 35 Figure 2 16 Setting the Bandwidth...
Page 142: ...132 SDG6000X User Manual Figure 2 95 Phase Deviation Interface...
Page 160: ...150 SDG6000X User Manual Figure 3 1 Generate a Sine Waveform...
Page 162: ...152 SDG6000X User Manual Figure 3 2 Generate a Square Waveform...
Page 175: ...SDG6000X User Manual 165 Figure 3 9 Generate an AM Modulation Waveform...
Page 177: ...SDG6000X User Manual 167 Figure 3 10 Generate a FM Modulation Waveform...
Page 179: ...SDG6000X User Manual 169 Figure 3 11 Generate a PM Modulation Waveform...
Page 181: ...SDG6000X User Manual 171 Figure 3 12 Generate a FSK Modulation Waveform...
Page 183: ...SDG6000X User Manual 173 Figure 3 13 Generate an ASK Modulation Waveform...
Page 185: ...SDG6000X User Manual 175 Figure 3 14 Generate a PSK Modulation Waveform...
Page 187: ...SDG6000X User Manual 177 Figure 3 15 Generate a PWM Modulation Waveform...
Page 189: ...SDG6000X User Manual 179 Figure 3 16 Generate a DSB AM Modulation Waveform...