127
Check the trigger mode on the trigger interface whether or not in the “normal” or
“single”, and check the trigger level is whether or not over the wave range. If yes,
please put the trigger level to the middle position or set the trigger mode to the
“Auto” position. In another hand, you could choose the “Auto” button to set up
automatically.
6. After the Acquisition is set to Averages or Display Persistence time is set too
long, the waveform refreshes slowly.
It is normal in these settings
7. The signal is displayed as ladder like waveform
(1)This phenomenon is normal. The time base maybe is too slow .you should
turn the horizontal SCALE knob to increase horizontal resolution to improve the
display.
(2) Maybe the display Type is set to “Vectors”, You could set it to Dots mode to
improve the display.