Appendix A: Specifications
All specification applies to 10X probe and All the SDS1000 Series Digital
Storage Oscilloscopes.
To verify that the oscilloscope meets specifications, the oscilloscope must
first meet the following conditions:
●
The oscilloscope must have been operating continuously for thirty minutes
within the specified operating temperature.
●
You must perform the Do Self Cal operation, accessible through the Utility
menu, if the operating temperature changes by more than 5° C.
●
The oscilloscope must be within the factory calibration interval
All specifications are guaranteed unless noted “typical.”
Specifications
Inputs
Input Coupling
AC, DC, GND
Input Impedance
1MΩ±2% || 16Pf±3Pf,
50
Ω
+/-2%(SDS1302CE/SDS1304CF,
SDS1202CE/SDS1204CF contain this function)
Maximum input voltage
400V (DC+AC PK-PK,1MΩ input impedance,X10),
CAT I
5Vrms;
+/-10V Peak for 1S pulse, 50% duty cycle(50
Ω
BNC)
(
SDS1302CE/SDS1304CF
,
SDS1202CE/SDS1204CF contain this function
)
Ch to Ch Isolation
(Both channels in same
V/div setting)
>100:1 at 150MHz: (SDS1302CE/SDS1304CF)
>100:1 at 100MHz: (SDS1202CE/SDS1204CF)
>100:1 at 70MHz (SDS1152CM)
>100:1 at 50MHz: (SDS1102C/CM/CE, SDS1104CF)
>100:1 at 30MHz: (SDS1062C/CM/CE, SDS1064CF)
>100:1 at 20MHz (SDS1042C)
>100:1 at 10MHz
(
SDS1022C
)
Probe attenuator
1X
、
10X
Probe attenuator Factors 1X
、
10X
、
100X
、
1000X