2 SDS1000CFL Service Manual
Waveform Intensity and Grid Brightness can be adjusted.
Menu display in the form of pop-up that in order to convenience users to
use it.
Multiple Language User Interface.
Support Multilingual online help system
Standard interface: Double USB Host, USB Device, LAN Port, Pass/Fail
Out
Specifications
These specifications apply to SDS1000CFL series Digital Storage
Oscilloscope. To verify that an oscilloscope meets specifications, it must first
meet the following conditions:
The oscilloscope must have been operating continuously for thirty minutes
within the specified operating temperature.
You must perform the Do Self Cal operation, accessible through the Utility
menu, if the operating temperature changes by more than 5 °C.
The oscilloscope must be within the factory calibration interval of one year.
All specifications are guaranteed unless noted “typical”.
Table 1-2 Oscilloscope specifications
Inputs
Input Coupling
AC, DC, GND
Input Impedance
(
1MΩ±2%
)
||
(
18Pf±3Pf
)
,50Ω±2%
Maximum input voltage
400Vpk, CAT I
Probe attenuation
1X, 10X
Probe attenuation Factors 1X, 5X, 10X, 50X, 100X, 500X, 1000X
Vertical System
Volts/Div Range
2mV/div~5V/div( in 1, 2, 5 sequence)
Channel Voltage
Offset Range
2mV~100mV: ±800mV
102 mV~5V: ±40V
Vertical Resolution
8 bit
The number of channel
2/4
70M (SDS1072CFL/SDS1074CFL)
100M (SDS1102CFL/SDS1104CFL)
200M (SDS1202CFL/SDS1204CFL)
Analog Bandwidth
300M (SDS1302CFL/SDS1304CFL)
Lower Frequency Limit
(AC Coupled)
≤10Hz at BNC