6 SDS1000CFL Service Manual
Prepare Information
Before doing performance verifying or procedure adjusting, you should master
the following operations to make the oscilloscope work in a good state or deal
with some simple functional problems. The following contents are included in
this chapter:
How to perform functional checks
How to operate four standard interface tests
How to use self-calibration routine
How to recall factory Default settings
Fore more detailed information about oscilloscope operation, please refer to
you Quick Guide for CFL.
Functional checking
This functional checking covers three kinds of checks, by which you can verify
whether the oscilloscope is working correctly.
Power-on Inspection
The normal operating voltage for SDS1000CFL series digital oscilloscope is in
range of 100-240VRMS with frequency from 45Hz to 440Hz. Connect the
power line to the socket on the rear panel of the oscilloscope.
Figure 2-1 Connect power line
Note: to avoid electric shock, make sure that the instrument is
correctly grounded to the earth before connecting AC power.
Press the power-on button located at the lower left corner of the front panel
and some keys will be lighted for about 6 seconds simultaneously until the boot
screen appears. Then the oscilloscope begins to perform its power-on tests
automatically, after that you can press the DEFAULT SETUP button to recall
the factory default settings.
The socket