14 SDS1000CFL Service Manual
USB Device Test
To test if the USB Device interface works normally connected with EasyScope
software.
Tools
:
● A SDS1000CFL oscilloscope
● A computer with USB interface
● A standard USB cable (Type AB)
● EasyScopeX software for SDS1000CFL
Steps
:
1. Set up EasyScopeX software in a computer and Install the driver step by
step following the instructions.
2. Connect the oscilloscope and the computer using an USB cable, and then
select “USBTMC” in “Back USB” menu under the UTILITY menu of the
oscilloscope.
3. Run EasyScopeX. Click “Add Device” at the upper left corner of the
EasyScope interface and immediately displays the “USBTMCDeviceDLG”
interface, then click the “Add” on the interface to complete the connection.