SIGLENT
SDS1000CFL Service Manual 19
After selecting the corresponding item as table above, you will see the
waveform displays as the following:
Figure 2-13 Pass/Fail Waveform
If the waveform is in the range that set under menu of Mask setting, it is
allowed to pass. Otherwise, it fails to pass.
4. Connect the Pass/Fail BNC terminal of CFL oscilloscope to a channel BNC
terminal of another oscilloscope.
5. If the Pass/Fail interface works normally, then you will see a pulse wave as
below displays on another oscilloscope by properly adjusting its time base
and voltage scale. Otherwise, there might be errors to the interface.
Figure 2-15 Pulse Waveform