20 SDS1000CFL Service Manual
Performance Test
This chapter mainly explains how to perform tests to make sure that the
oscilloscope meets the performance specifications. For accurate test results,
please let the test requirement and the oscilloscope warm up 30 minutes
before testing.
Here are the required equipments for the test:
Table 3-1 Test equipments
Description
Critical Specifications
Example
DC Voltage Source
-6mv~30v, ±0.1% accuracy
Sine Wave Generator
10Hz~300MHz,
±0.1% amplitude accuracy
FLUKE 9500B
Calibrator
BNC Cable
BNC(m) to BNC(m),
About 1m long
50Ω Feedthrough
Termination
50Ω BNC(f) to BNC(m)
Here is the connecting between FLUKE 9500B and the oscilloscope under
test:
Picture 3-1 Connecting test instruments