SIGLENT
24 SDS1000CFL Service Manual
300MHz if you are testing a SDS1302/1304CFL
Table 3-4 Bandwidth test data
Volts/Div
Input frequency
Vpp measured
10Hz
1.2V
100Hz
1KHz
10KHz
100KHz
1MHz
10MHz
20MHz
…
200mV/div
bandwidth
10Hz
600mV
100Hz
1KHz
10KHz
100KHz
1MHz
10MHz
20MHz
…
100mV/div
bandwidth
7. Every time you set to a frequency, it is better for you to adjust the
Sec/Div to a proper value thus to display a perfect waveform.
8. The V
pp
measurement of the waveform will decrease as the
frequency increases, check the frequency, which is defined as the
bandwidth, until the V
pp
measurement is closest to 0.848v.
9. Disconnect the test connection.
10. Check the other channels in the same way as step 1 to 9.
For test at 100mV/div, please perform the same operation as that at 500mV/div,
except for the initial amplitude and amplitude at bandwidth listed below.
Table 3-4 Amplitude at bandwidth
Volts/Div
Initial
amplitude
Amplitude
at bandwidth
Initial
frequency
200mV/div
1.2V
0.707
×
1.2V=0.848v
10 Hz
100mV/div
600mV
0.707
×
600mV=424mV
10 Hz