SDS1000CFL Service Manual 25
To verify Time Base Accuracy
This test checks the time base accuracy of one channel. In the test, both the
impedance of FLUKE and the oscilloscope are required to set to 50Ω. For
detailed information about adding 50Ω feedthrough terminator with different
types of CFL oscilloscope at the end of BNC cable connected to FLUKE 9500B,
please refer to corresponding information in bandwidth verification above.
Time Base Accuracy: Freq/10M < 50ppm
Steps:
1. Connect selected channel of the oscilloscope to the Sine Wave
Generator of FLUKE 9500B.
2. Set the frequency of 9500B to 10MHz.
3. Set the oscilloscope Volts/Div to 500mv/div.
4. Set the oscilloscope Sec/Div to 1ms/div.
5. Press MEASURE button on the front panel of the oscilloscope to
display Freq measurement.
6. Calculate Freq/10M to see if the value is within the range of 50ppm,
which means correct time base accuracy.
7. Disconnect the test connection.