SIGLENT
26 SDS1000CFL Service Manual
To verify Trigger Sensitivity
To test trigger sensitivity at the frequency of 100KHz, 10MHz and bandwidth.
Table 3-5 Trigger sensitivity data
Volts/Div
Frequency
Trigger Range
500mv/div
100KHz
500mVpp
500mv/div
10MHz
500mVpp
500mv/div
bandwidth
750mVpp
Steps:
1. connect the selected channel of the oscilloscope to the Sine Wave
Generator of FLUKE 9500B.
2. Set the oscilloscope Volts/Div to 500mv/div.
3. Set the oscilloscope Sec/Div to 5us/div.
4. Set the frequency of the Sine Wave Generator to 100KHz.
5. Set the amplitude of the Sine Wave Generator to 500mVpp, so that
the Vpp measurement of the oscilloscope is approximately 500mv.
6. Press Trigger LEVEL button to move the waveform to the screen
center.
7. Adjust the Trigger LEVEL within the waveform range to check if
there is stable trigger.
8. Set the oscilloscope Sec/Div to 50ns/div.
9. Set the frequency of the Sine Wave Generator to 10MHz.
10. Set the amplitude of the Sine Wave Generator to 500mVpp, so that
the Vpp measurement of the oscilloscope is approximately 500mv.
11. Press Trigger LEVEL button to move the waveform to the screen
center.
12. Adjust the Trigger LEVEL within the waveform range to check if
there is stable trigger.
13. Set the oscilloscope Sec/Div to 5ns/div.
14. Set the frequency of the Sine Wave Generator to:
70MHz if you are testing a SDS1072/1074DFL
100MHz if you are testing a SDS1102/1104DFL
200MHz if you are testing a SDS1202/1204DFL
300MHz if you are testing a SDS1302/1304DFL
15. Set the amplitude of the Sine Wave Generator to 750mVpp, so that